![vacuubrand MZ 1C Скачать руководство пользователя страница 53](http://html1.mh-extra.com/html/vacuubrand/mz-1c/mz-1c_instructions-for-use-manual_3209466053.webp)
page 53 of 92
➨
Air and pumped media might react inside the pump or
at the outlet of the pump and form hazardous or explo-
sive mixtures, when you use air rather than inert gas
for the gas ballast. This constitutes a risk of significant
damage to equipment and/or facilities, a risk of per-
sonal injury or even loss of life.
+
Make sure that air/gas intake through the gas ballast
valve can never lead to hazardous, explosive or other-
wise dangerous mixtures. If in doubt, use inert gas.
To reduce condensation in the pump, do not pump vapor
before the pump has reached its operating temperature.
Open the gas ballast valve when pumping condensable
vapors. Turn gas ballast cap to open valve.
NOTICE
gas ballast
(open)
For
condensable vapors
(water vapor, sol-
vents, etc.):
- The gas ballast valve is open if the arrow
on the gas ballast cap is pointing away from
the pump (
MD 1C / MD 1C VARIO-SP)
, re-
spectively towards the pump (
MD 1C + AK
+ EK
) or away from the inlet (
MZ 1C
) (see
figures).
- With gas ballast valve open, the ultimate
vacuum will be reduced.
- Use inert gas for gas ballast to avoid the
formation of explosive mixtures.
Attention
:
maximum supply pressure of inert gas: 17.5
psi (1.2 bar) absolute.
- Close the gas ballast valve by turning the
cap 180°.
In case of low boiling solvents (when the for-
mation of condensate is unlikely), the use of
gas ballast might be unnecessary. Operating
the pump without gas ballast increases the
solvent recovery rate at the exhaust waste va-
por condenser (EK).
gas ballast (open)
MD 1C + AK + EK
MD 1C / MD 1C VARIO-SP
gas ballast
(open)
MZ 1C
Содержание MZ 1C
Страница 30: ...page 30 of 92 MD 1C MD 1C C US only pump with dual voltage motor 7 10 11 6 9 8 5 4 3 2 3 2 3 2 ...
Страница 31: ...page 31 of 92 MZ 1C 3 2 7 10 11 6 9 8 ...
Страница 32: ...page 32 of 92 MD 1C AK EK 10 3 6 1 8 2 9 7 15 16 17 13 11 8 14 19 18 ...
Страница 33: ...page 33 of 92 MD 1C VARIO SP 7 10 6 12 11 8 Ensure sufficient venting of the pump ...
Страница 63: ...page 63 of 92 Checking diaphragms and valves 1 2 3 MD 1C AK EK 10 mm 2 5 mm MD 1C AK EK MD 1C AK EK 5 1x ...
Страница 64: ...page 64 of 92 4 3A 5 MD 1C size 2 2x 1 17 mm 14 mm max 1 4 ...
Страница 66: ...page 66 of 92 6 7 8 4 6x Clean ...
Страница 67: ...page 67 of 92 9 Replacing the diaphragm 10 11 20696828 20696838 ...
Страница 68: ...page 68 of 92 12 1 2 3 13 14 Pay attention to number of washers ...
Страница 69: ...page 69 of 92 15 16 17 Pay attention to number of washers ...
Страница 70: ...page 70 of 92 18 1 2 19 20 4x Replacing the valves and assembling the pump heads ...
Страница 71: ...page 71 of 92 21 22 23 ...
Страница 73: ...page 73 of 92 27 28 1 2 28A size 2 2x 17 mm 10 mm MD 1C AK EK 5 1x MD 1C ...
Страница 74: ...page 74 of 92 29 30 31 Assembling the connecting hose MD 1C AK EK 2x ...
Страница 77: ...page 77 of 92 5 6 4 20638821 20638821 ...
Страница 89: ...page 89 of 92 ...
Страница 90: ...page 90 of 92 ...
Страница 91: ...page 91 of 92 ...