![vacuubrand MZ 1C Скачать руководство пользователя страница 15](http://html1.mh-extra.com/html/vacuubrand/mz-1c/mz-1c_instructions-for-use-manual_3209466015.webp)
page 15 of 92
• Connect hoses gas tight at inlet and outlet of the pump.
• Ensure that no foreign objects can be drawn into the
pump.
• Check the power source and the pump’s rating plate
to be sure that the power source and the equipment
match in voltage, phase, and frequency.
• On pumps with a dual-voltage motor (100-120V / 200-
230V; 50/60 Hz), check that the voltage selection
switch is set correctly. Do not change the setting of the
voltage selection switch while the pump is connected
to AC power. Unplug the pump before setting the volt-
age selection switch.
Note
: If the pump is switched on
with wrong voltage selection, the motor may be dam-
aged!
• Ensure that no liquids can flow on or into the pump mo
-
tor when assembling or disassembling vacuum lines at
the pump. Risk of corrosion and/or short circuit!
• Ensure that the
coolant outlet pipe
is always free and
that it cannot get blocked. If installing an optional cool-
ant valve, it must always be in the supply line of the
exhaust waste vapor condenser.
Make sure ventilation is adequate to maintain recommend-
ed operating temperature. Keep a minimum distance of
2 in (5 cm) between the cooling fan and surrounding items
(e.g., housing, walls, etc.), or else install an external auto-
matic ventilation system. Clean fan guard grill if necessary
to avoid a reduction of ventilation.
Use only hoses at the inlet and outlet of the pump with
an inner diameter at least as large as the diameter of the
pump’s tubing (to avoid overpressure at the outlet, and
reduction of pumping speed at the inlet).
Allow the equipment to equilibrate to ambient temperature
if you bring it from cold environment into a room prior to
NOTICE
Содержание MZ 1C
Страница 30: ...page 30 of 92 MD 1C MD 1C C US only pump with dual voltage motor 7 10 11 6 9 8 5 4 3 2 3 2 3 2 ...
Страница 31: ...page 31 of 92 MZ 1C 3 2 7 10 11 6 9 8 ...
Страница 32: ...page 32 of 92 MD 1C AK EK 10 3 6 1 8 2 9 7 15 16 17 13 11 8 14 19 18 ...
Страница 33: ...page 33 of 92 MD 1C VARIO SP 7 10 6 12 11 8 Ensure sufficient venting of the pump ...
Страница 63: ...page 63 of 92 Checking diaphragms and valves 1 2 3 MD 1C AK EK 10 mm 2 5 mm MD 1C AK EK MD 1C AK EK 5 1x ...
Страница 64: ...page 64 of 92 4 3A 5 MD 1C size 2 2x 1 17 mm 14 mm max 1 4 ...
Страница 66: ...page 66 of 92 6 7 8 4 6x Clean ...
Страница 67: ...page 67 of 92 9 Replacing the diaphragm 10 11 20696828 20696838 ...
Страница 68: ...page 68 of 92 12 1 2 3 13 14 Pay attention to number of washers ...
Страница 69: ...page 69 of 92 15 16 17 Pay attention to number of washers ...
Страница 70: ...page 70 of 92 18 1 2 19 20 4x Replacing the valves and assembling the pump heads ...
Страница 71: ...page 71 of 92 21 22 23 ...
Страница 73: ...page 73 of 92 27 28 1 2 28A size 2 2x 17 mm 10 mm MD 1C AK EK 5 1x MD 1C ...
Страница 74: ...page 74 of 92 29 30 31 Assembling the connecting hose MD 1C AK EK 2x ...
Страница 77: ...page 77 of 92 5 6 4 20638821 20638821 ...
Страница 89: ...page 89 of 92 ...
Страница 90: ...page 90 of 92 ...
Страница 91: ...page 91 of 92 ...