7.
Replace the used O‑ring.
8.
Clean the pressure distributor if
necessary.
9
Top
Bottom
10
9.
Place the new pressure relief
valve on the clean surface.
Ensure the pressure relief
valve is positioned correctly on
the pressure distributor.
10.
Position the suction distributor
with screws and wind in the
screw fittings until hand‑tight;
Phillips screwdriver size 2.
11
12
11.
Push the molded hoses back
into place on the hose noz‑
zles.
12.
Secure the hose clips on the
hose nozzles, e. g., with flat
nose pliers.