28
20901115_12+16C_VARIO_sel_V1.0_130618
Installation and connection
Check installation
conditions
>
Please note that a
dia‑
phragm pump can weigh
approx. 25 kg
. We recom‑
mend using a lifting aid.
>
Lift the unit out of the pack‑
aging by the side handles.
4.2 Installation
Note
Condensate can damage the electronics.
A large temperature difference between the storage location
and the installation location can cause condensation.
>
>
After goods receipt or storage, allow your vacuum device to
acclimatize for at least 3‑4 hours before initial use.
Check installation conditions
The device is acclimatized.
Ambient conditions have been observed and are within the limi‑
tation of use.
The pump must have a stable and secure base without addi‑
tional mechanical contact apart from the pump feet.
Example
Lift out the
diaphragm pump
(shown with
vapor condenser
accessory)
Содержание MV 10C VARIO select