Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check
carefully the validity of this document with respect to his product. manual-no.: 995652  / 08/10/2009

During operation:

Max. ambient temperature

: 40 °C

Make sure ventilation is adequate if pump is installed in a housing or if ambient
temperature is elevated.

If pump is installed in altitudes of more than 1000 m above mean sea level check
compatibility with applicable safety requirements, e. g. IEC 60034 (motor may over-
heat due to insufficient cooling).

The motor is shut down by a 

thermal cutout

 in the winding.

Attention: Manual reset is necessary. Switch off the pump or isolate the equipment
from mains. Wait approx. five minutes before restarting the pump.

Attention

: In case of 

supply voltage below 100V

, the lock of the cutout might be

restricted and the pump might restart on its own after sufficient cooling down. Take
suitable precautions, if an automatic restart of the pump may lead to a critical
dangerous situation.

Due to the high compression ratio of the pumps, pressure at the outlet port might be
generated being higher than the max. permitted pressure compatible with the mechani-
cal stability of the system.

Do not start pump if 

pressure difference between inlet and outlet port exceeds

max. 1 bar

. Attempts to start pump at higher difference may cause blockade and

damage of the motor.

Check compatibility with 

max. permitted pressure at outlet 

and

 max. pressure

difference between inlet and outlet ports

.

Prevent internal condensation, transfer of liquids or dust. The diaphragm and valves
will be damaged, if liquids are pumped in significant amounts.

A warm up period (approx. 15 min.) is required to ensure that the rated ultimate total
pressure and pumping speed are attained.

Shutdown:

Has the pump been exposed to condensate?

Allow the pump to continue to run at atmospheric pressure for a few minutes.

Has the pump been exposed to media which may damage the pump materials or form
deposits?

Check and clean pump heads if necessary.

Содержание ME 2

Страница 1: ...completely and unchanged It is strictly the users responsibility to check carefully the validity of this document with respect to his product manual no 995652 08 10 2009 Diaphragm pumps ME 2 ME 4 MZ...

Страница 2: ...r pump VACUUBRAND diaphragm pumps are the result of many years of experience in construction and practical operation of these pumps combined with the latest results in material and manufacturing techn...

Страница 3: ...ct to his product manual no 995652 08 10 2009 Attention Important notes Not permitted Misuse may cause damage Caution Hot surface Isolate equipment from mains Note Contents Safety information 4 Techni...

Страница 4: ...protected electrical supply and a suitable earth point Failure to connect the motor to ground may result in deadly electrical shock The supply cable may be fitted with a moulded European IEC plug or...

Страница 5: ...pheres If pumping different substances purge the pump with inert gas prior to changing the pumped media in order to pump out residues and to avoid reactions of the pumped substances with each other wi...

Страница 6: ...ts The A weighed emission sound pressure level of the pump does not exceed 70 dB A Measurement according to EN ISO 2151 2004 and EN ISO 3744 1995 with standard silencer or exhaust tube at outlet Ensur...

Страница 7: ...o m t a o n n o i t a r e p o g n i r u d t a s n e d n o c f o n o i t a r a p e s r e w o p r o t o M W 0 2 1 0 8 1 0 8 1 0 8 1 n o i t c e t o r p r o t o M t u o t u c l a m r e h t w a r d t n e...

Страница 8: ...M P F 2 Z M 2 E M e l z z o n e s o h T B P 4 E M e l z z o n e s o h l e e t s s s e l n i a t s D 2 Z M e g n a l f l l a m s l e e t s s s e l n i a t s r e c n e l i s E P A P e s o h E P g n i t...

Страница 9: ...of this document with respect to his product manual no 995652 08 10 2009 ME 2 with dual voltage motor 100 120 200 230V ME 2 outlet with silencer mains cable on off switch outlet with silencer inlet p...

Страница 10: ...ct manual no 995652 08 10 2009 ME 4 MZ 2 MZ 2D fig MZ 2D inlet outlet with silencer pump rating plate motor rating plate on off switch mains cable inlet MZ 2 hose nozzle outlet pump rating plate motor...

Страница 11: ...igid pipelines Insert elastic hoses or flexible elements as couplings between the pump and rigid pipes Use a suitable valve to isolate the pump from the vacuum system if you need to allow the pump to...

Страница 12: ...oling down Take suitable precautions if an automatic restart of the pump may lead to a critical dangerous situation Due to the high compression ratio of the pumps pressure at the outlet port might be...

Страница 13: ...roduct manual no 995652 08 10 2009 ME 2 MZ 2 Small flange DN 16 662590 Clamping ring DN 10 16 aluminium 660000 Centring ring DN 10 16 NBR 660190 Hose nozzle for MZ 2D 639758 ME 4 Small flange DN 16 63...

Страница 14: ...840 Pressure adjustment device outlet 696841 Installation vacuum and or pressure adjustment device at ME 2 MZ 2 Assemble control valve first then align and fasten gauge Installation vacuum adjustment...

Страница 15: ...a t l o v e v e i h c a t o n s e o d p m u P r o e r u s s e r p l a t o t e t a m i t l u d e e p s g n i p m u p l a m r o n y l t c e r r o c t o n g n i r g n i r e t n e C e h t n i k a e l r o...

Страница 16: ...ump is exposed to corrosive gases or vapour or in case of deposits mainte nance should be carried out frequently Regular maintenance will improve the lifetime of the pump and also protect both man and...

Страница 17: ...ound hole clamping disc with countersunk head screw diaphragm with square hole clamping disc with square head screw Use open ended wrench remove the screw in fittings together with the hose from all p...

Страница 18: ...mp head Smaller number of washers The pump will not attain ulti mate vacuum More washers Clamping disc will hit head cover noise or even blockade of the pump Insert diaphragm support disc and new diap...

Страница 19: ...eccentric bushing bring connecting rod into up per turning point position Max stroke of the rod Pay attention that the diaphragm stays positioned centrally so that it will become clamped uniformly bet...

Страница 20: ...p chamber By measuring the pressure at the inlet port of the individual head Use a suitable vacuum gauge e g DVR 2 cat no 68 29 02 make sure that it is correctly calibrated and measure the pres sure a...

Страница 21: ...ry product returned for repair We submit quotations only on request and always at the customer s expense If an order is given the costs incurred are offset from the costs for repair or from the pur ch...

Страница 22: ...for the returned product that neither toxic corrosive biologically active explosive radio active nor contamination dangerous in any way has occurred the product is free of dangerous substances the oi...

Страница 23: ...e dispositif d sign ci dessus est conforme aux prescriptions de base des directives EU applicables et indiqu s en ci que concerne conception dessin et mod le vendu par nous m mes Cette d claration ces...

Страница 24: ...ature is only intended to inform our customer The validity of general empirical values and results obtained under test conditions for specific applications depend on a number of factors beyond our con...

Отзывы: