General information
☞
Read and comply with this manual before installing or operating the equip
-
ment.
☞
Transport the pump at the provided handle.
Remove all packing material, remove the product from its packing-box, remove the
protective covers from the inlet and outlet ports and keep, inspect the equipment.
If the equipment is damaged, notify the supplier and the carrier in writing within three
days; state the item number of the product together with the order number and the
supplier’s invoice number. Retain all packing material for inspection.
Do not use the equipment if it is damaged.
If the equipment is not used immediately, replace the protective covers. Store the
equipment in suitable conditions.
Intended use
☞
Use the equipment for the intended use only according to the ATEX conformity, e.
g. for pumping gas atmosphere without aerosol in a range form 0.1 to 1000 mbar
and a maximum inlet pressure (dynamic pressure) and outlet pressure of 1100
mbar (absolute).
☞
The pump and all system parts must not be used on humans or animals.
☞
Prevent any part of the human body from coming into contact with vacuum.
☞
Ensure that the individual components are only connected, combined and oper-
ated according to their design and as indicated in the instructions for use.
☞
Comply with notes on correct vacuum and electrical connections, see section
”Use and operation”.
The pumps are marked with ”X” (according to EN 13463-1), i. e. restrictions of
the operation conditions:
• The pumps are designated for a low degree of mechanical stress and have
to be installed in a way so that they can not be damaged from outside.
Pumping units with glass attachment parts have to be installed protected
against shocks from the outside and against glass splinters in the event of
breaking (implosion). Attention: The joint clips at the glass attachment parts
are not grounded. Avoid charge.
• The pumps are designated for an ambient and gas inlet temperature during
operation of +10 to +40°C. Never exceed these ambient and gas inlet tem
-
peratures.
Setting up and installing the equipment
Before system start-up of the pump or the complete system the user has to evaluate
in its entirety the explosion risks and takes adequate measures in application of the
principle of risk assessment.
NOTICE
NOTICE