![vacuubrand MD 4C NT + 2 AK Скачать руководство пользователя страница 59](http://html.mh-extra.com/html/vacuubrand/md-4c-nt-2-ak/md-4c-nt-2-ak_instructions-for-use-manual_847963059.webp)
page 64 of 75
If the pump does not achieve the ultimate vacuum:
- Whenever the diaphragms and valves have been replaced, a break-in
period of several hours is required before the pump achieves its ulti-
mate vacuum.
- In case of an unusual noise, switch off pump immediately and check
clamping disc positions.
change after the break-in period:
Check hose connectors at pump heads for leaks. If necessary recheck
valve seats and pump chambers.
Checking the ultimate vacuum
After any intervention at the equipment (e.g., repair / maintenance)
the ultimate vacuum of the pump has to be checked. Only if the pump
enough to ensure that no explosive atmospheres will occur in the inte-
rior of the equipment.