valve (see ”Accessories”, pg. 48)
in the supply line
of the exhaust vapor condenser to save water and re-
duce the risk of water spill.
by blocked or kinked coolant hoses).
In case of
condensation
: Check the liquid level in both
catchpots (9) during operation. Check the liquid level in
both catchpots regularly. Do not allow the catchpots to
level sensor (see ”Accessories”, pg. 48) for monitoring,
if necessary (VACUUBRAND controller CVC 3000 or VNC
2 is required).
The maximum liquid level is at approximately 80% of the
catchpots.
Permissible range of coolant temperature at the exhaust
waste vapor condenser: 5 °F to 68 °F (-15°C to +20°C)
Check hose connections prior to starting operation of the
cooling system.
Check coolant hoses regularly during operation.
Removing the catchpots:
Catchpot at outlet:
Remove joint clip. Remove catchpot and drain conden-
sate.
Catchpot at inlet:
Admit air or inert gas (via the pump inlet) to restore at-
mospheric pressure in the catchpot before attempting
removal. Remove joint clip. Remove catchpot and drain
condensate.
Reattach drained catchpots.
NOTICE
NOTICE
Содержание MD 4C NT + 2 AK
Страница 25: ...page 30 of 75 ME 4C NT 1 5 4 7 6 8 2 ME 2C NT 1 5 4 7 6 8 2 ...
Страница 27: ...page 32 of 75 ME 8C NT MD 4C NT 1 5 3 4 7 6 2 8 8 5 4 7 6 1 ...
Страница 50: ...page 55 of 75 Fittings and tubing of the different pump models ME 4C NT MZ 2C NT ME 8C NT MD 4C NT ...
Страница 67: ...page 72 of 75 tive mark Licensed Test mark on the pump rating plate ...
Страница 68: ...page 73 of 75 China RoHS ...
Страница 69: ...page 74 of 75 ...