![vacuubrand MD 4C NT + 2 AK Скачать руководство пользователя страница 37](http://html.mh-extra.com/html/vacuubrand/md-4c-nt-2-ak/md-4c-nt-2-ak_instructions-for-use-manual_847963037.webp)
page 42 of 75
(e.g., coolant valve).
by blocked or squeezed coolant hoses).
of the exhaust waste vapor condenser.
During operation
Vent and dispose of
potentially dangerous gases or
vapors
at the outlet of the pump appropriately.
Due to the high compression ratio, the pump might
generate overpressure at the outlet. Check pressure
compatibility with system components (e.g., exhaust
tubing or exhaust valve) at the outlet. Ensure that the
pump outlet is neither blocked nor restricted.
Maximum ambient temperature
: 104 °F (40 °C)
Check the maximum temperatures, if installing the
pump in a cabinet or a housing. Make sure ventilation
is adequate, especially if the ambient temperature is
elevated.
ft (2000 m) above mean sea level, check compatibility
with applicable safety requirements, and adopt suit-
able measures. There is a risk of the motor overheat-
maximally permitted
pressure
at outlet and the
maximum pressure dif-
ference
between inlet and outlet ports.
Do not start the pump if the
pressure difference between
inlet and outlet ports exceeds max. 16.0 psi (1.1 bar)
.
Attempts to start the pump at higher pressure difference
may cause stalling and damage of the motor.
If pumping condensable vapors (water vapor, solvents,
NOTICE
Содержание MD 4C NT + 2 AK
Страница 25: ...page 30 of 75 ME 4C NT 1 5 4 7 6 8 2 ME 2C NT 1 5 4 7 6 8 2 ...
Страница 27: ...page 32 of 75 ME 8C NT MD 4C NT 1 5 3 4 7 6 2 8 8 5 4 7 6 1 ...
Страница 50: ...page 55 of 75 Fittings and tubing of the different pump models ME 4C NT MZ 2C NT ME 8C NT MD 4C NT ...
Страница 67: ...page 72 of 75 tive mark Licensed Test mark on the pump rating plate ...
Страница 68: ...page 73 of 75 China RoHS ...
Страница 69: ...page 74 of 75 ...