Installation and connection
4.4.3
4.4.3 Venting connection (option)
Venting connection (option)
DANGER
DANGER
Risk of explosion by venting with air.
Depending on the application, venting can cause explo
‑
sive mixtures to form or other hazardous situations to
arise.
>
Never vent processes with air which could form an
explosive mixture.
>
In the case of flammable substances, use
only inert gas for venting, e. g., nitrogen
(max. 1.2 bar/900 Torr abs.).
The diaphragm pump and controller have
no
no
direct ventilation
connection. You can connect different venting valves, e. g., the
supplied
VACUU·SELECT
®
Sensor
with integral venting valve.
(b)
The ventilation connection
(b)
for a
VACUU·SELECT
®
Sensor
is described
below.
Alternatively you can use a larger valve,
e. g., a
VB M-B
(#20674217) for faster
venting.
Vent with ambient air
1
For venting
(b)
with ambient air, nothing needs to be connected to
the sensor
(a)
.
Vent with inert gas – connect venting valve
1
Required connection material:
Hose for hose nozzle, e. g., sili
‑
cone tube 4/5 mm
>
Attach hose
(c)
to the connection of venting valve
(b)
and con‑
nect inert gas (max. 1.2 bar/900 Torr, abs.).
5
Venting valve with hose for venting with inert gas
2
.
1 Not applicable to sensors without an integrated venting valve .
2 Avoid overpressure .
(c)
(b)
(a)
(a)
(b)
(a)
(a)
VACUU·SELECT
®
Sensor with venting
valve
Содержание MD 12C NT VARIO o.C.
Страница 26: ...26 20901544_EN_Mx10C 12C 16C_NT Vsel_V1 2_260922 Product description...
Страница 46: ...46 20901544_EN_Mx10C 12C 16C_NT Vsel_V1 2_260922 Commissioning operation...
Страница 90: ...90 20901544_EN_Mx10C 12C 16C_NT Vsel_V1 2_260922...
Страница 91: ...91 20901544_EN_Mx10C 12C 16C_NT Vsel_V1 2_260922...