6
Leak
finder
®
Initial Setup,
continued
4
.
Air Compressor
Air pressure between 3.4 to 17 bar
(50 ~ 250 PSIG)
It is very important to keep the
Leak
finder
® nitrogen generator assembly free of
oil, water and other contaminants, in order to maintain a high nitrogen purity
and long service life. In order to do that, you need to be sure and maintain a
clean filter system on your air supply into the
Leak
finder
®. In addition, install
the Filter Assembly Kit WVRP-104, or equivalent.
Under normal use, we recommend you replace the filter elements inside Filter
Assembly Kit
WVRP-104
once a year and use only approved Filter Elements Kit
Part No. WVRP-103, as noted in Optional Accessories page in this manual. Turn
filter bowl ¼ turn in counter-clock direction to gain access to the filter elements.
IMPORTANT NOTE:
In order for the moisture and contaminants that collect in the
filter bowls to drain, air must be disconnected and re-
connected to the machine’s
Air Fitting. So do not leave air source to the machine connected for days at a time,
or the filters will never drain, which can damage the nitrogen generator system. An
occasional disconnect and reconnect to air is sufficient.