![Vacc-Safe VSI-120 Скачать руководство пользователя страница 25](http://html2.mh-extra.com/html/vacc-safe/vsi-120/vsi-120_operating-manual_4788690025.webp)
VACC-SAFE CO
2
Incubator
25
write
40004
Set CO
2
density CO
2
Int
Unit:0.1%
Read and
write
40005
Audible alarm
Int
0: Null
1: activate
Read and
write
Input register
Register address
(DEC.)
Explanation
Data type
Remark
Read or write
30001
Current chamber
temperature
Int
Unit:0.01
℃
Read
30002
Current door
temperature
Int
Unit:0.01
℃
Read
30003
Current sterilization
temperature
Int
Unit:0.01
℃
Read
30005
Current CO
2
density
Int
Unit
:
0.01%
Read
30006
Open door time
Int
Unit
:
1 second,
MAX:240
Read
30007
Sterilization
remaining Time
Int
Unit:1 second
Read
30008
Temperature deviates
more than ±2
℃
Int
0
:
Null
1
:
Activate
Read
30009
Chamber
temperature sensor
disconnection alarm
Int
0
:
Null
1
:
Activate
Read
30010
Chamber
temperature exceeds
60.5
℃
alarm
Int
0
:
Null
1
:
Activate
Read
30013
CO
2
density deviates
more than ±1%
Int
0
:
Null
1
:
Activate
Read
300014
CO
2
sensor
disconnection
alarm
CO
2
Int
0
:
Null
1
:
Activate
Read