3. Prerequisite and configuration
3.1. Prerequisite
Inspection of Boxes
When you have received the instrument which is packed on pallet, inspect the box carefully for any
damages that may have caused any damages to product during shipping. Please report any damage
to the carrier or to your local VACC-SAFE distributor.
LOCATION
The incubator is designed to operate at temperature 5
℃
above ambient, and recommended to
operate at minimum ambient 15
℃
Maximum Room Temperature is 32
℃
.
To avoid place for use this incubator is as below.
Near Heater or Freezer (if it may generate heat and affect temperature control of incubator)
Near Equipment generating heat or cold air to incubator.
Directly Sunlight Exposed to incubator
Uneven ground or table head
The place where is being vibrated
Too narrow to use lift handle (at side of bottom) and power cable of incubator.
Cleaning before use
Before conducting cell culture. It’s recommended to clean up entire chamber and shelves, water tray
by using at least 70% Ethanol mixed of 30% distilled water and soft clothes.
Inserting shelves
Shelves are mounted onto the shelf racks in such a way that the edge of shelves which is slightly bent
up about 8mm goes to the back of the chamber until it is too close to the back wall. Basically, 3 shelve
are provided. Insert the shelves from top to bottom.
Level the incubator by adjusting feet. Place a small level on the second shelf of the incubator adjust
the level feet until the incubator is level and stable.
After inserting the shelves, place humidity tray in the bottom of chamber if humidification is required
for your application.