20
(& CAL/SET) which directed as CLEAN MODE.
At the same time long press two buttons to enter the dry heat sterilization process
The display should indicate the following:
Figure 2: Display of the incubator indicating sterilization cycle
“CLEAN” is shown in temperature display and 600 is shown in CO₂ LED Window. CLEAN means the
HOT AIR Decontamination program is now activated. The number 600 displayed in CO₂ LED is the time
counted down from 600 minutes (10 hours). This show how many minutes are left until the program
finishes.
If the temperature has reached at pre-set value
;
(Last programmed set point before hot air
decontamination), the display will no longer show “CLEAN”. It will in turn to the normal
temperature and CO
2
values. This indicates the incubator is ready for normal use again.
6.8 UV decontamination
UV Lamp ON /OFF Switch
UV
UV Lamp is located near circulation fan and above ceiling of chamber. It decontaminates circulation
fan and in-air circulated from fan. Most light is blocked by ceiling.