Electrical installation
WARNING
:
• Observe the specifications of the examination certificate valid in your country.
• Observe the laws and regulations valid in your country for electrical installations in potentially explosive
atmospheres.
• Refer to Hazardous Area Installation on page 31 if applicable.
• Ensure that the available power supply complies with the power supply specified on the product nameplate
and specified in the examination certificate valid in your country.
• Shipping plugs in the cable inlets must be replaced by suitable screwtype glands or dummy plugs, which are
appropriately certified for transmitters with explosion-proof protection.
• For CE installations, use a cable with a braided metallic shield (or armoured cable where applicable).
• The lid must not be opened in wet locations while the unit is powered. (A wet location is a location where
water or another conductive fluid may be present and is likely to increase the risk of electric shock.)
Notes
:
• Lay PROFIBUS PA cable separately from power cable with voltages greater than 60 V.
• Avoid locating the unit near large electrical equipment wherever possible.
• Connect the cable shield to earth (for example, to the housing by means of a metallic screwed gland).
WARNING
:
• All field wiring must have insulation suitable for at least 250 V.
• Only qualified personnel are authorized to install and operate this equipment in accordance with established
safety practices and standards.
• The Protective Earth Terminal indicated by
must be connected to reliable ground. In case of
non-metallic vessels, the external earth wire should be connected to an earthed component which is
earthed near the vessel.
• All wiring must be done by qualified personnel in accordance with all governing regulations.
• The equipment must be protected by a 15A fuse or circuit breaker in the building installation.
• A circuit breaker or switch in the building installation, marked as a disconnect switch, shall be in close
proximity to the equipment and within easy reach of the operator.
• Use shielded cable, wire gauge 20 AWG to 14 AWG (0.5 mm
2
to 2.0 mm
2
). For CE installations use a cable
with a braided metallic shield (or armoured cable where applicable).
• Maximum working voltage between adjacent relay contacts is 250 V.
• Relay contact terminals are for use with equipment which has no accessible live parts and wiring which has
insulation suitable for at least 250 V.