![UWT NivoRadar NR 7100 Series Скачать руководство пользователя страница 43](http://html1.mh-extra.com/html/uwt/nivoradar-nr-7100-series/nivoradar-nr-7100-series_technical-information-instruction-manual_1066137043.webp)
NivoRadar
®
Continuous level measurement
Series NR 7100
Technical information / Instruction manual
Diagnostics and servicing
page 43
Fault description
Cause
Rectification
Measured value fluctuates
around 10 … 20 %
Level
time
0
Various echoes from an uneven medi-
um surface, e.g. an extraction funnel
Check parameter "Material Type" and
adapt, if necessary
Reflections from the medium surface
via the vessel wall (deflection)
Optimize installation position and sen-
sor orientation
How to proceed if a repair is necessary
You can find an instrument return form as well as detailed informa-
tion about the procedure in the download area of our homepage.
By doing this you help us carry out the repair quickly and without
having to call back for needed information.
Proceed as follows in case of repair:
•
Print and fill out one form per instrument
•
Clean the instrument and pack it damage-proof
•
Attach the completed form and, if need be, also a safety data
sheet outside on the packaging
•
Ask the agency serving you to get the address for the return
shipment. You can find the agency on our homepage.