UV
222
Industrial
Technical & Installation Manual
16
If during the duty cycle, something/someone comes closer to the unit than the
programmed safety off distance, the Product will automatically stop, and it will resume the
activity 10 sec. after the security distance is cleared.
If the information programmed into the Product when installed is different than the use/
occupancy of the room, UV Medico A/S cannot be held liable for any injuries or misuse of
these units.
If there are changes in the use, occupancy or any other altenrations that has an influence
on the risks of exposure, the installer needs to be contacted so the Product is repro
-
grammed.
11) Testing the Product.
Right click on the photo of the Product
A
and a pop up of the test mode will show up.
Here you can quickly light up the Product and test the signal LED.
When you activate the 222 LightSource
B
, it will turn on the Product with
no safety
. This
is only to test the source, (the box “Enable“ will be automatically ticked).
When you go to LED Mode you have 3 options:
•
Normal Mode: Work as the Product is programmed.
•
LED Off: Turns the LED off.
•
LED RGBW Off: LED test cycle.
When you untick the “Enable” box it goes back to normal configuration/use.
When configuration is done, disconnect the USB cable and screw the cap on gently,
withtout force.
A
B