![Utah Medical Products FINESSE + Скачать руководство пользователя страница 36](http://html.mh-extra.com/html/utah-medical-products/finesse/finesse_service-manual_1064775036.webp)
32
Service Manual
Pass
Verify smoke evacuation flow rate switch changes motor speed.
Done Adjust volume control so that cut and coag audible tones can be clearly heard over the
smoke evacuation motor set at the ‘high’ flow rate position.
Output Calibration
Set cut mode output to "99", mode to "C", and electrosurgical analyzer load to 500
Ω
. Line
voltage MUST be 114.0-116.0 VAC for FIN-110 and 228.0-232.0 VAC for FIN-220. Activate
system and adjust R334 until output is 97W. NOTE: Output may vary several watts due
to feedback activity.
Pass
_____ (95W-99W)
Check system output power at the following displayed settings. All measurements should be
made through a 500
Ω
load. Tolerances are listed in parentheses. Do not activate system for
more than 10 seconds for each reading, and allow at least 20 seconds between readings.
Individual output calibration is not adjustable.
Pass Cut
30: _____ (25-35W) 60: _____ (51-69W) 90: _____ (77-103W)
Pass Blend 1
30: _____ (25-35W) 60: _____ (51-69W) 90: _____ (77-103W)
Pass Blend 2
30: _____ (25-35W) 60: _____ (51-69W) 90: _____ (77-103W)
Pass Blend 3
30: _____ (25-35W) 60: _____ (51-69W) 90: _____ (77-103W)
Pass Coag
30: _____ (25-35W) 60: _____ (51-69W) 75: _____ (64-86W)
HF Leakage Currents
Set cut display to “99” and coag display to “75”. Establish the HF leakage current path by
connecting a non-inductive 200
Ω
load between the active electrode and mains ground. Leave
dispersive connection open. Activate system to check active HF leakage current:
Active HF Leakage
Pass Cut _____ (<150mA)
Pass Coag _____ (<150mA)
Connect non-inductive 200
Ω
load resistor between dispersive electrode and mains ground. Leave
active connection open. Activate system to check dispersive HF leakage current:
Dispersive HF Leakage
Pass Cut _____ (<150mA)
Pass Coag _____ (<150mA)
Содержание FINESSE +
Страница 12: ...8 FINESSE Service Manual ...
Страница 18: ...14 FINESSE Service Manual ...
Страница 24: ...20 FINESSE Service Manual ...
Страница 34: ...30 FINESSE Service Manual ...
Страница 42: ...38 FINESSE Service Manual Mains Power and System Wiring Diagram ...
Страница 43: ...FINESSE Service Manual 39 ...
Страница 45: ...FINESSE Service Manual 41 RF Power Amplifier PCB Drawing ...
Страница 46: ...42 FINESSE Service Manual Logic PCB Drawing 90738 B ...
Страница 47: ...FINESSE Service Manual 43 Display PCB Drawing 90737 B ...
Страница 48: ...44 FINESSE Service Manual Power Supply PCB Drawing 90740 ...
Страница 52: ...48 FINESSE Service Manual Notes ...
Страница 53: ...FINESSE Service Manual 49 ...