26
Service Manual
Smoke Evacuation Duration
The smoke evacuation system will begin running immediately on activation of the
electrosurgery module and remain running five seconds after the output power is deactivated.
Disposable Filter Cartridge
The FINESSE Filter Pack, catalog number ESU-501, consists of a pleated HEPA paper filter
followed by a compartment containing activated charcoal.
The FINESSE Filter Pack, to ensure total elimination of odors between procedures, should be
replaced on a daily basis. However, the three-stage filtration system has been tested to
effectively remove odors and particles for up to 15 electrosurgical procedures. Over a period
of time the external disposable filter pack can be a source of odor and possible viral
contamination. Therefore, it is recommended that the external filter pack be changed every
day or after 15 procedures if more than 15 procedures are performed in a single day.
Third-Stage Internal Filter
The Internal Filter, item number SSE-500, is a pleated ULPA filter element. Annual
replacement is recommended.
Particle Removal Efficiency
Spherical particles with a mean diameter of 0.1 microns were removed with a minimum
efficiency of 99.999%, as determined by laboratory tests via a latex particle challenge.
Vacuum Motor Exhaust Filter
The smoke evacuation motor is equipped with a HEPA grade particulate filter. This
filter protects the 's internal electronics from any particulate that may be generated
by the motor. This filter should never need cleaning or changing. However, if the smoke
evacuation system flow seems degraded after replacing all other filtration components,
contact Utah Medical Products for information.
Environmental Specifications
Operational Environment
Temperature:
50
˚
F to 104˚F (10˚C to 40˚C)
Humidity:
10% to 93%
Pressure:
700 hPa to 1060 hPa (10.15 PSI to 15.36 PSI)
Storage and Transport Environment
Temperature:
-40
˚
F to 158˚F (-40˚C to 70˚C)
Humidity:
10% to 93%
Pressure:
500 hPa to 1060 hPa (7.25 PSI to 15.36 PSI)
After transport and/or storage at conditions outside of the operating environment range, allow
time for the system to reach the operating environment — typically one hour is
sufficient.
Содержание FINESSE +
Страница 12: ...8 FINESSE Service Manual ...
Страница 18: ...14 FINESSE Service Manual ...
Страница 24: ...20 FINESSE Service Manual ...
Страница 34: ...30 FINESSE Service Manual ...
Страница 42: ...38 FINESSE Service Manual Mains Power and System Wiring Diagram ...
Страница 43: ...FINESSE Service Manual 39 ...
Страница 45: ...FINESSE Service Manual 41 RF Power Amplifier PCB Drawing ...
Страница 46: ...42 FINESSE Service Manual Logic PCB Drawing 90738 B ...
Страница 47: ...FINESSE Service Manual 43 Display PCB Drawing 90737 B ...
Страница 48: ...44 FINESSE Service Manual Power Supply PCB Drawing 90740 ...
Страница 52: ...48 FINESSE Service Manual Notes ...
Страница 53: ...FINESSE Service Manual 49 ...