14.
REDOX MEASUREMENT AMPLIFIER
14.1
PARAMETER MENU (STANDARD REDOX AMPLIFIER)
TEXT
DESCRIPTION
POSSIBLE SETTINGS
DISPLAY
CODE
Prevention of unauthorized
Acces code 904
access to following menus
SP = mV
Setpoint (contact)
All values in the selected
range (F menu)
Al = mV
HI alarm
Settings from limit value
min to top value of
measuring range
Al = mV
LO alarm
Settings from bottom value
of measuring range to limit
value max
Ald (min)
Limit time delay
0...60 min
Alh mV
Limit hysteresis for metering
1...50 digits of display
and alarm contact
Al N.O.
Alarm relay
N.O. = normally open
N.C. = normally closed
--- switched off
0-20 mA
Selection of mA output
0...20/4...20/0...5/0...10 mA
--- switched off
INCREASE/
Direction of control
Control upwards or
DECREASE
downwards
M. PUMP
Control element selection menu
Solenoid pump, continuous
output, electric positioner,
metering pump
NOTE:
For Redox-amplifier the option "Controller" may be ordered. For this option the
following parameter menus are valid.
47
Содержание DEPOLOX 4
Страница 10: ......
Страница 41: ......
Страница 54: ......
Страница 55: ......
Страница 56: ......
Страница 60: ......
Страница 73: ......
Страница 74: ......
Страница 75: ......
Страница 76: ......
Страница 77: ......
Страница 78: ......
Страница 79: ......
Страница 80: ......
Страница 81: ......
Страница 82: ......
Страница 83: ......
Страница 84: ......
Страница 85: ......
Страница 86: ......
Страница 87: ......
Страница 88: ......
Страница 89: ......
Страница 90: ......
Страница 91: ......
Страница 92: ......
Страница 93: ......
Страница 94: ......
Страница 95: ......
Страница 96: ......
Страница 97: ......
Страница 98: ......
Страница 99: ......
Страница 100: ......
Страница 101: ......
Страница 102: ......
Страница 103: ......
Страница 104: ......
Страница 105: ......
Страница 106: ......
Страница 107: ......
Страница 108: ......
Страница 109: ......
Страница 110: ......
Страница 111: ......
Страница 112: ......
Страница 113: ......