MPACT ENDOSCOPIC INSUFFLATOR
®
SYSTEM Operator’s Manual
Page 37
Table 3
Guidance and
manufacturer’s declaration – electromagnetic immunity
The CO
2
MPACT ENDOSCOPIC INSUFFLATOR
®
SYSTEM is intended for use in the electromagnetic environment specified
below. The customer or the user of the CO
2
MPACT ENDOSCOPIC INSUFFLATOR
®
SYSTEM should assure that it is used in
such an environment.
Immunity Test
IEC 60601 test level
Compliance
level
Electromagnetic environment
–guidance
Portable and mobile RF communications equipment should be
used no closer to any part of the CO
2
MPACT ENDOSCOPIC
INSUFFLATOR
®
SYSTEM, including cables, than the
recommended separation distance calculated from the equation
applicable to the frequency of the transmitter.
Recommended separation distance
IEC 61000-4-6
3 Vrms
3 Vrms
d
= 1.2
P
Conducted RF
150 kHz to 80 MHz
6 Vrms
6 Vrms
ISM
IEC 61000-4-3
3 V/m
3 V/m
d
= 1.2
P
80 MHz to 800 MHz
Radiated RF
80 MHz to 2.7 GHz
d
= 2.3
P
800 MHz to 2.7 GHz
where
P
is the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer and
d
is the
recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey
a
, should be less than the compliance
level in each frequency range
b
.
Interference may occur in the vicinity of equipment marked with
the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular / cordless) telephones and land mobile radios,
amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the
electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the
measured field strength in the location in which the CO
2
MPACT ENDOSCOPIC INSUFFLATOR
®
SYSTEM is used exceeds
the applicable RF compliance level above, the CO
2
MPACT ENDOSCOPIC INSUFFLATOR
®
SYSTEM should be observed to
verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or
relocating the CO
2
MPACT ENDOSCOPIC INSUFFLATOR
®
SYSTEM.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m