CO
2
MPACT ENDOSCOPIC INSUFFLATOR
®
SYSTEM Operator’s Manual
Page 33
Condition
Possible Causes
Field Checkout
My CO
2
cylinder is
prematurely emptying.
What could be the problem?
The valve on the CO
2
cylinder left open all of the
time.
High Pressure Hose from
CO
2
cylinder is not installed
properly.
We recommend attaching your CO
2
MPACT ENDOSCOPIC
INSUFFLATOR
®
SYSTEM to C, D, or E size CO
2
cylinders. For days when the CO
2
MPACT ENDOSCOPIC
INSUFFLATOR
®
SYSTEM is scheduled for continuous use,
the CO
2
cylinder valve should be opened at the start of the
shift and closed at the end of the shift.
The pneumatic design of the CO
2
MPACT ENDOSCOPIC
INSUFFLATOR
®
SYSTEM is not intended to provide a full
time gas seal from the CO
2
cylinder while the device is not
being used or in storage. Should the CO
2
cylinder valve
inadvertently be left open, there is a high likelihood that the
CO
2
cylinder contents will gradually empty over several
days.
If after opening and closing the CO
2
cylinder valve with your
procedure schedule, you still encounter problems with the
CO
2
cylinder prematurely emptying, there is the possibility
that the High Pressure Hose from the CO
2
cylinder to the
CO
2
MPACT ENDOSCOPIC INSUFFLATOR
®
SYSTEM
unit is leaking.
Please inspect the High Pressure Hose in accordance with
the following procedure: Close Valve on CO
2
tank. Using
an adjustable or open ended ¾" wrench, detach the High
Pressure Hose from the CO
2
MPACT ENDOSCOPIC
INSUFFLATOR
®
SYSTEM rear panel. Inspect the tapered
surface of the free standing nipple on the CO
2
MPACT
ENDOSCOPIC INSUFFLATOR
®
SYSTEM and the interior
of the High Pressure Hose. This is the sealing surface.
Verify that it is free of any debris or contaminant. Also,
verify that the surface is smooth and free of any nicks or
distortion.
Repeat this inspection procedure at the identical tapered gas
connection at the CO
2
cylinder valve Yoke.
Taper
Taper