MPACT ENDOSCOPIC INSUFFLATOR
®
SYSTEM Operator’s Manual
Page 18
9.2 PREPARATION TEST
There is always the possibility that delicate equipment can be damaged in
transportation or storage. Therefore it is important to verify proper operation of the
unit before use.
1. After power is applied to the unit, verify that the front panel lights are enabled.
2. To verify that flow control is functioning properly, depress the START/STOP
button. When pressed, the adjacent LED light will illuminate green, and gas may
be heard exiting from the unit. If gas does not begin flowing, verify that the CO
2
cylinder valve is in the open position (see Section 9.1,
Power On and Gas Supply
Indicator
, above).
3. To prevent wasting gas, when the START/STOP button is pressed initially, an
internal timer will start. Depending on the TIMER setting selection, the flow of
CO
2
will automatically stop after 30 minutes, 60 minutes, or 120 minutes. The
TIMER can be pre-selected by the user depending on the anticipated length of
the procedure. After the pre-selected time has elapsed, the unit automatically
returns to STOP mode. Thereafter, subsequent presses of the START/STOP
button will resume the delivery of CO
2
for an additional 30, 60, or 120 minutes,
depending on the pre-selected time. However, the flow can be stopped at any
time using the START/STOP button.
NOTE: The device will default to 60 Minutes when powered on.
4. If the unit does not perform properly, do not use. Inspect the unit using the
Troubleshooting guide (Section 13.0) before returning for service.
9.3 TUBING SET CONNECTION
Insert the connector on the Tubing Set to the Gas Output connection on the unit’s
front panel. It is important to use only US Endoscopy manufactured high flow tubing
sets, that includes a ≤0.1
micron hydrophobic filter, designed to provide optimum
performance. Always inspect every Tubing Set to make sure there are no signs of
damage. If such a condition exists, do not use the Tubing Set. See Tubing Set
instructions for use.