CO
2
MPACT ENDOSCOPIC INSUFFLATOR
®
SYSTEM Operator’s Manual
Page 16
8.4 CONNECTING TO A CO
2
GAS PIPELINE ADAPTER
1.
The optional DISS High Pressure Hose (REF 710604) is required for connection
to the CO
2
gas pipeline.
2.
Attach the 7/16” female connector (see Figure 8.2) on the CO
2
Central Supply
Hose to the male CO
2
gas connector on the rear panel of the CO
2
MPACT
ENDOSCOPIC INSUFFLATOR
®
SYSTEM
and securely tighten it.
3.
Connect the DISS connector (see Figure 8.3) on the CO
2
Central Supply Hose to
the CO
2
gas pipeline hose and tighten the adapter manually until the position
where the connector is stopped.
Figure 8.2
Figure 8.3
Before each use the following procedures or inspections should be performed:
•
Visually inspect the CO
2
MPACT ENDOSCOPIC INSUFFLATOR
®
SYSTEM
for external signs of damage.
•
Inspect the electrical connections. Do not use if inspection reveals damage.
•
Before connecting the power cord to the wall outlet, make sure the main power
switch is off and that the voltage is correct. Inspect the connection to the CO
2
supply tank, to assure it is intact and tight.