Z12 ZOOM STEREO MICROSCOPE
UNITRON
73 Mall Drive, Commack, NY 11725 • 631-543-2000 • www.unitronusa.com
16
TROUBLESHOOTING
(continued)
MECHANICAL PART
PROBLEM
CAUSE
SOLUTION
The coarse adjustment knob is
too difficult to rotate
The tension adjustment ring is tightened too
much
Loosen it
The image goes out of focus
during observation or the stage
drops by itself
The tension adjustment collar is too loose
Tighten it
ELECTRICAL SYSTEM
PROBLEM
CAUSE
SOLUTION
The lamp doesn’t light
No power to the lamp
Check the power cord is connected correctly
The lamp is not installed correctly
Install it correctly
The lamp burns out
Replace the lamp
The light intensity is not
enough
The light intensity control knob is not set
properly
Adjust the light intensity control knob
The wrong lamp is used
Replace the lamp with the correct one
The lamp keeps burning out
The wrong lamp is used
Replace the lamp with the correct one
The lamp flickers or the
brightness is not stable
The lamp will burn out soon
Replace the lamp
The power supply doesn’t connect well
Connect the power supply correctly