background image

IMT120 15 

4

Type 820E

To adjust the switch:

 

Remove the set point adjustment knob and clear cover 

from the indicating enclosure (square) to access the pointer 

mechanism for adjustment.

❷ 

Connect test lights or listen for the individual switch clicks 

to indicate switch operation. 

❸ 

Move the set point adjustment pointer (red) up the scale 

past the indicating pointer (black). This permits checking the 

set point by moving the actuating lever upward with a finger, 

or tool, simulating thermal assembly movement. Note switch 

operation on the test lights.

❹ 

Loosen adjustment pointer mounting screw “A” and move the 

adjustment pointer (red) until it aligns with the indicating pointer 

(black) (see Figure 5). When the switch clicks re-tighten the screw. 

Type 822E

NOTE: 

Type 822E has a single set point adjustment-

knob and adjustment pointer (red) for both switches. 

Both switches are standardly factory set together 

within 5% of the adjustable set point range. Turning 

the set point adjustment knob adjusts both switches 

simultaneously, maintaining a fixed separation 

between them. To adjust the switches:

 

Remove the explosion-proof enclosure cover to access the 

switches for adjustment.

❷ 

Remove the set point adjustment knob and clear cover 

from the indicating enclosure (square) to access the pointer 

mechanism for adjustment. 

❸ 

Connect test lights or listen for the individual switch clicks 

to indicate switch operation. 

 

Move the adjustment pointer (red) up the scale past the 

indicating pointer (black), actuating both switches, by moving 

the actuating lever upward with a finger, or a tool, simulating 

thermal assembly movement (see Figure 5). Note switch 

operation on the test lights. 

❺ 

The set point of switch #1 is determined by the adjustment 

pointer (red). The set point of switch #2 may be adjusted by 

up to 25% of the range span below the switch #1 set point 

by turning adjustment “D” with a 5/16” open end wrench 

(clockwise) to lower the set point. See page 6 for 822E dual 

switch adjustment. The set point of switch #2 should never be 

set above that of switch #1.

 To align the switch #1 adjustment pointer (red) with the 

indicating pointer (black), loosen the adjustment pointer mounting 

screw  “A” and align the adjustment pointer (red) with the indicating 

pointer (black), then re-tighten screw “A” (see Figure 5).

NOTE: Indicating Pointer Deflection (820E, 822E)

The indicating pointers will read slightly low when 

the bulb temperature is 15° F above the controller 

setting. This deflection is normal and repeatable 

(approximately 0.5% of scale range on single switch 

models) and is due to the transference of the switching 

mechanism load to the thermal system. It can be 

measured by moving the setting pointer from the high 

to the low end of the scale and observing the resultant 

indicating pointer deflection.

NOTE: 

To adjust the Indicating Pointer (820E, 822E): 

 

Use an accurate test thermometer such as a thermocouple 

with its probe mounted directly to the center of the sensing bulb. 

❷ 

Before making any adjustments, allow process temperature 

to stabilize; i.e., successive on-off cycles repeated.

 Correct any difference between the indicating pointer 

(black) and the test thermometer by holding the compensator 

with a 5/16” wrench while turning the zero adjustment “C” on 

the thermal assembly with a second 5/16” wrench (see Figure 

5) until the indicating pointer (black) reads the same value 

as the test thermometer. Turning the zero adjustment “C” 

clockwise lowers indicated reading. 

 

Compare the process temperature with the set point 

adjustment pointer (red). 

❺ 

Loosen the adjustment pointer mounting screw “A” to align 

set point adjustment pointer (red) with the indicating pointer 

(black), then re-tighten screw “A”.

Correction of Capillary (820E, 822E)

If the length of capillary immersed in the process 

differs from the amount immersed (6”, see Mount

-

ing Bulb & Capillary in Part I - Installation) at the 

factory calibration bath, a calibration shift will 

occur. 

The error may be corrected as follows: 

 Move pointer to the highest temperature setting. Note 

indicating scale reading with the head and sensor at room 

temperature. 

❷ 

Loosen the two thermal assembly mounting screws.

❸ 

Re-position the housing index against the calibration on 

the instrument case (or skeleton casting) at a rate of 1 division 

line per capillary length listed in Figure 6. Move to the left 

if capillary is to be added to the process, or to the right if 

capillary is to be removed from process. 

Figure 5: Indicating 

Enclosure Internals

CAPILLARY

INDICATING 

POINTER 

(BLACK)

THERMAL 

ASSEMBLY 

MOUNTING 

SCREWS (2)

THERMAL 

ASSEMBLY 

HOUSING

INDEX

ADJUSTMENT 

POINTER (RED)

ADJUSTMENT 

POINTER 

MOUNTING 

SCREW 

(ADJUSTMENT “A”)

ZERO 

ADJUSTMENT 

“C” 5/16 HEX

COMPENSATOR

ACTIVATING  

LEVER

Содержание 120 Series

Страница 1: ...ARE AS FOLLOWS 120 s Activation Plunger to adjustment screw hole gap joints 1 140 in 28 96 mm min length by 0 0039 in 0 099 mm max annular gap Plunger Guide to enclosure through threaded joints minim...

Страница 2: ...ION STEM HEX SEE DIMENSIONAL DRAWINGS DO NOT TIGHTEN BY TURNING THE ENCLOSURE AS THIS WILL DAMAGE THE SENSOR AND WEAKEN WELDED JOINTS The conduit connection must be properly sealed potted for horizont...

Страница 3: ...d turning the calibrating screw with another 3 16 open end wrench see figure 3 Turn clockwise to increase setting or counter clockwise to decrease setting Types B121 and E121 Remove adjustment cover A...

Страница 4: ...then re tighten screw A see Figure 5 NOTE Indicating Pointer Deflection 820E 822E The indicating pointers will read slightly low when the bulb temperature is 15 F above the controller setting This de...

Страница 5: ...B122 9 84 250 0 Immersion stem 2BS 8BS 9 31 236 6 Bulb capillary 13242 13322 10 0 254 0 Immersion stem Freeze protection 13273 13321 9 31 236 6 Bulb capillary Heat tracing Model Number Range Cap Lengt...

Страница 6: ...ELL INSTALLATIONS ii Except model 13273 13321 bulb OD is 1 4 6 35 mm 6 3 8 161 9 mm 3 3 16 81 0 mm 3 3 16 81 0 mm 7 177 8 mm 6 3 8 161 9 mm 14 11 16 373 1 mm 9 228 6 mm 7 8 22 2 mm 1 3 16 30 2 mm 3 4...

Страница 7: ...Y INSTALLED Il est imp ratif d utiliser des raccords d tanch it antid flagrants pour les fils lectriques Les types C120 F120 820E et 822E poss dent deux ouvertures de conduits lectriques de 3 4 NPT l...

Страница 8: ...disassembled by anyone other than authorized Seller s representatives EXCEPT FOR THE LIMITED WARRANTY OF REPAIR AND REPLACEMENT STATED ABOVE SELLER DISCLAIMS ALL WARRANTIES WHATSOEVER WITH RESPECT TO...

Отзывы: