Calibration of U137, U235 & 237 Series Weight Indicators.

Date:141027   B01950   1(10)

Unisystem  AB
Bergebyv. 24

Tel.:

 0560-140 55

www.unisystem .se

685 34 TORSBY (Säte)

Fax:

 0560-101 25

info@ unisystem .se

Index.

Page

1 Coarse gain, offset and response time. Keyboard functions at calibration. Calibration sequence.

Page

2 Calibration sequence.

Page

4 Calibration sequence.

Page

5 Calibration sequence. U2375 with tilt sensor. Restrictions for verified scales.

Page

6 Setpoints. Programmable serial output codes. Serial input.

Page

7 Transducer connector. U237 connector. U137 RS232 connector. Internal connectors.

Page

8 U2375 tilt transducer connector. Optional connectors.

Page

9 Precautions when installing! Flow measurement. Pictures.

Page 10 Pictures.

Coarse sensitivity (gain), offset and response time.

In order to make coarse adjustments of sensitivity, offset and low pass filter, the bottom panel must be removed

and the boards taken out. Be very careful with the flat cable to the front panel! All switches and jum pers are situated
on the main board behind the display board, under the shield. Normally these adjustments are not necessary.

The accuracy of the settings is better than 0.5% .

Input sensitivity, corresponding to 93700(0) AD-units:

Response tim e to 0.1% , optional resistors:

S2:

1,3

3

1

2

on

J15-16-17

Standard

J15-16-17

2.673

1.663

1.021

0.634

mV/V

410ms

R10, R15 removed.

2x4.99M

290ms

2x10M

2360ms

The offset is set by:

2x1.74M

206ms

2x4.99M

1180ms

JP10

3-6

5-6

4-5

1-4

2-5

on

2x715k

120ms

2x2.74M

648ms

JP11 off

0.000

0.314

0.410

0.508

0.821

mV/V

2x383k

75ms

2x1.74M

410ms

JP11 on -0.052

0.262

0.358

0.456

0.769

mV/V

2x169k

37ms

2x715k

200ms

The response time to 0.01% error is twice the time above.
The 3dB frequency in Hz is 1/(2.2 x the time above in seconds)

The following values are set at delivery: Sensitivity 2.673mV/V, offset -0.052mV/V and response time 410 ms.

 Lower input sensitivity with no offset may be received by:
 3.424mV/V for JP10 off, S2:1,3 on. Connect R11:12  to R11:13. Disconnect U11:8.
 4.272mV/V for JP10 off, S2:1,3 on. Connect R11:6 to 11 and 12.. U12:1 to R11:10. Disconnect U11:8. (U2396)

U235

 has a fixed offset of -0.069mV/V.

The sensitivity is 2.673m V/V but may be changed to 1.663m V/V (W5 cut off) or 1.021m V/V (W6 cut off).
There is a 3-pole filter with low ringing. The response time is 1.6 seconds or 0.5 seconds with S2:1, 2 and 3 on.

Keyboard functions at calibration.

The software calibration is performed via the serial input or the keyboard (U235 and older U2373 must have a

separate keyboard, U2395). The functions of the keys at calibration are as follows:

{ZERO}

Steps forward one position in the calibration sequence. Previous step is stored.

{TARE}

Steps back one position in the calibration sequence. Previous step is stored.

{NET/GROSS}

Selects the digit position of data. In sign position, the NET indicator blinks.
Switches between increment and decrement in Cs1 to 22. (Cs = Calibration step).

{COUNT}

Increments or decrements the digit value and changes sign.

{F} {ZERO}

Cs23-25. Gets the signals AD-value. {F} immediately after gets the old value back.
Cs31, 32. Sets the displayed AD-value to zero e.g. at 0° tilt for easier calculation.

{F} {TARE}

Extrapolates the AD-value in Cs25 from Cs23 and Cs24 when max weight is missing

{PRINT/TEST}

Prints the calibration sequence and program information.

{F} {F}

Returns to the AD-value (before Cs1). Skips change after {F} {ZERO} or {TARE}.

{F} {PRINT/TEST} {ZERO}

Enters preset values marked * below, when the AD-value is displayed e.g. after {F} {F}.

Calibration sequence.

The calibration switch S1 is located on the bottom panel. W hen S1 is switched on, the software calibration mode is

entered and the 6 digit internal AD-value is displayed. U137 and U235 only presents the 5 most significant digits for
the AD-value.

In the following calibration sequence, the functions and calibration may be changed.
W hen S1 is switched off, the indicator enters weighing mode. Restart the indicator to get all changes.

The calibration sequence is also entered without S1 by {F}{ZERO}, 3 successive times (from 001127). Cs 01 to 15

may be changed but not the calibration. The sequence is left by {F}{F}.

The display shows XX  NN. XX is Cs(Calibration step) for 01 - 22, 28 and 29. NN is the chosen function code.
= code cannot be used together with any other = code in the same Cs.
+ code can be combined with any other + code and one = code in the Cs.

  Add all + codes and the = code to get the number NN to enter in the Cs.

* means preset data according to the above. New separate indicators are delivered with these values.

Steps 23 - 27 and 30 - 32 have more sub steps according to below.

Not used steps, may be used for special functions, and NN = 0 must be chosen.

Содержание U137 Series

Страница 1: ...21mV V W6 cut off There is a 3 pole filter with low ringing The response time is 1 6 seconds or 0 5 seconds with S2 1 2 and 3 on Keyboard functions at calibration The software calibration is performed...

Страница 2: ...RS422 or in Cs1 3 4 RS422 is addressed by ASCII 124 1 to 14 CR No one is addressed by and character other than 1 to 9 CR is recommended 15 Always addressed Used for single RS4225 unit without address...

Страница 3: ...3 4 ranges 1 at 10 20 50 2 at 10 25 50 5 at 10 20 40 4 5 ranges 1 at 5 10 25 50 2 at 4 10 20 40 5 at 5 10 20 50 5 6 ranges 1 at 2 4 10 20 40 2 at 2 5 10 20 50 5 at 2 5 5 10 25 50 6 Not used 7 Calibrat...

Страница 4: ...change in AD value 25 Third calibration point 00000 Display data 3 Must be even number 000000 Internal AD value Cs25 is not used when this value is lower than that in Cs24 Full AD converter range is...

Страница 5: ...both tilts and used in the formula The calibration can not be left with S1 when displaying angle These errors are caused by different angle between tilt sensor and load cell s and by the side force se...

Страница 6: ...mm month and dd day Option 190 Time as hh mm ss where hh are hours mm minutes and ss seconds Option 191 Date as 20yy mm dd where yy are the two last digits of the year mm month and dd day Option 192 W...

Страница 7: ...tation voltage 10V 1 However U2379 5V 1 Connect GND to metal hood of D sub J3 U237 series Power interface connector 9p D sub male B01910 1 2 3 4 5 6 7 8 9 0V Return for J3 2 Power V 0VD VB V after fus...

Страница 8: ...25mA SYNC CK SYNC D1 SYNC D2 TARE PT RET PRINT TEST DTR Connect to metal hood of D sub RS232 B01910 U137 has inverted TTL output Use J3 for RS232 RS232 U137 has inverted TTL output Use J3 for RS232 R...

Страница 9: ...be made in separate grounded metal boxes Beware of high temperature sources as the signal is very sensitive to temperature gradients The indicator must be well protected against high humidity dust and...

Страница 10: ...ht Indicators Date 141027 B01950 10 10 Unisystem AB Bergebyv 24 Tel 0560 140 55 www unisystem se 685 34 TORSBY S te Fax 0560 101 25 info unisystem se Old display U2370 U2371 U2372 U2373 U2375 U2379 U2...

Отзывы: