Calibration of U137, U235 & 237 Series Weight Indicators.

Date:141027   B01950   7(10)

Additional serial input functions at calibration or program m able serial output sequence:

F 50 P

The indicator switches to and from transm itting the calibration sequence help text when stepping in
the calibration sequence.

F n Z  (or A)

Jumps to calibration step n in the sequence. Two consecutive jumps are allowed, then it restarts.

m Z  (or A)

Enters value m  in the step and goes to the following or preceding step. At the programmable serial
output sequence, the values are also stored in the EEPROM.

V(86)

Enters programmable serial output sequence. Corresponds to {ZERO} and {TARE} simultaneously.

RS422.

 The serial input is activated by 

|

(ASCII 124) and 

1-14

 and 

[CR]

 (ASCII 13). Address 1-14 in Cs04.

All inputs are deactivated by 

|  [CR]

Note!

 Never com municate to other equipm ent with an indicator addressed and do not use the character 

|

 else.

Connectors.

J2:

Transducer connector. 9p D–sub female.

1
2
3
4
5
6
7
8
9

+SIGNAL
–SIGNAL
 0VA  Analog zero.
+SENSE
+EXC  Excitation.
GND  Ground/shield.
+5VA
–SENSE
–EXC  Excitation.

NOTE! The body of the transducer (load cell) must be grounded!

The input amplifier is easily destroyed, when an electrostatic discharge occurs on
an ungrounded transducer, with power on the indicator. There is a great risk,
when an unm ounted load cell is tested separately, as there normally is no inter-
nal ground connection. More regarding this below under precautions.
Use shielded cable, preferably double shielded, and filter D-sub.
For non symmetric load cells, +excitation to signal voltage must be higher than
signal to –excitation! If not, change all polarities.
Excitation voltage 10V ±1% . However U2379 5V±1% .
Connect GND to metal hood of D-sub.

J3:

U237 series. Power/interface connector. 9p D–sub male. B01910.

1
2
3
4
5
6
7
8
9

0V Return for J3:2
Power +V
0VD
+VB (+V after fuse)
TD
Ground/shield.
RD
CTS
DTR

J3:2. 11 - 26V. opt. 29V. Always connect direct to the battery, when there is a

pp

risk of parallel inductive loads. Ripple >0.1V  increases variation (slow jitter).
U2372. Charge battery on +V via series diode. +VB is battery voltage after the
internal fuse for e.g. printer. When 0VD is used for return, +VB is switched on/off
by the indicator. Max 0,5A, 1A interm ittent.
Note! The asynchronous output pin configuration does not correspond to RS232,
but the signal level is standard. Option U1731, RS232 with 25p D-sub.
No asynchronous signals in U2373 and U2379.
Use shielded cables.
Connect GND to metal hood of D-sub.

J3:

U137 series. RS232 interface connector. 9p D–sub male. B01910. Option U23890: RS422.

1
2
3
4
5
6
7
8
9

J4:1
RD
TD
DTR
0VD, J4:2
RS422 Rx A. J4:3.
RS422 Rx B. J4:6.
RS422 Tx A. J4:5.
RS422 Tx B. J4:4.

J4 is for optional signals.

RS422 is option. Full duplex.
Use shielded cable and terminating resistors at the cable ends.

Connect GND to metal hood of D-sub.

J1:

Internal I/O connector except U2373.

1
2
3
4
5
6
7
8
9

10
11
12
13
14
15
16

GND
0VD
+5V
TARE
NET/GROSS
COUNT
F
PRINT/TEST
ZERO
(TD)
(DTR)
(RD)
(CTS)
SYNC D2
SYNC D1
SYNC CK

Connect to 0VD for function. R=10k
Connect to 0VD for function. R=10k
Connect to 0VD for function. R=10k
Connect to 0VD for function. R=10k
Connect to 0VD for function. R=10k
Connect to 0VD for function. R=10k
U237, RS232 level. U137, inverted, 5V CMOS level.
U237, RS232 level. U137, inverted, 5V CMOS level.
U237, RS232 level. U137, inverted, 5V CMOS level.
U237, RS232 level. U137, inverted, 5V CMOS level.
Synchronous data 2. HCMOS level. B01900,
Synchronous data 1. HCMOS level.
Synchronous clock. HCMOS level.

Содержание U137 Series

Страница 1: ...21mV V W6 cut off There is a 3 pole filter with low ringing The response time is 1 6 seconds or 0 5 seconds with S2 1 2 and 3 on Keyboard functions at calibration The software calibration is performed...

Страница 2: ...RS422 or in Cs1 3 4 RS422 is addressed by ASCII 124 1 to 14 CR No one is addressed by and character other than 1 to 9 CR is recommended 15 Always addressed Used for single RS4225 unit without address...

Страница 3: ...3 4 ranges 1 at 10 20 50 2 at 10 25 50 5 at 10 20 40 4 5 ranges 1 at 5 10 25 50 2 at 4 10 20 40 5 at 5 10 20 50 5 6 ranges 1 at 2 4 10 20 40 2 at 2 5 10 20 50 5 at 2 5 5 10 25 50 6 Not used 7 Calibrat...

Страница 4: ...change in AD value 25 Third calibration point 00000 Display data 3 Must be even number 000000 Internal AD value Cs25 is not used when this value is lower than that in Cs24 Full AD converter range is...

Страница 5: ...both tilts and used in the formula The calibration can not be left with S1 when displaying angle These errors are caused by different angle between tilt sensor and load cell s and by the side force se...

Страница 6: ...mm month and dd day Option 190 Time as hh mm ss where hh are hours mm minutes and ss seconds Option 191 Date as 20yy mm dd where yy are the two last digits of the year mm month and dd day Option 192 W...

Страница 7: ...tation voltage 10V 1 However U2379 5V 1 Connect GND to metal hood of D sub J3 U237 series Power interface connector 9p D sub male B01910 1 2 3 4 5 6 7 8 9 0V Return for J3 2 Power V 0VD VB V after fus...

Страница 8: ...25mA SYNC CK SYNC D1 SYNC D2 TARE PT RET PRINT TEST DTR Connect to metal hood of D sub RS232 B01910 U137 has inverted TTL output Use J3 for RS232 RS232 U137 has inverted TTL output Use J3 for RS232 R...

Страница 9: ...be made in separate grounded metal boxes Beware of high temperature sources as the signal is very sensitive to temperature gradients The indicator must be well protected against high humidity dust and...

Страница 10: ...ht Indicators Date 141027 B01950 10 10 Unisystem AB Bergebyv 24 Tel 0560 140 55 www unisystem se 685 34 TORSBY S te Fax 0560 101 25 info unisystem se Old display U2370 U2371 U2372 U2373 U2375 U2379 U2...

Отзывы: