Calibration of U137, U235 & 237 Series Weight Indicators.

Date:141027   B01950   9(10)

Precautions when installing!

The transducer signal inputs are easily damaged and must not be touched or left open with power on.

0VD is connected to 0VA. Both must only be connected to GND once, normally in the power supply at W1.

Shielded cables must always be used, J2 preferably double shielded, with twisted, shielded pairs. The shields must

always be connected at both ends. It is normally not done inside the transducer. The cable hoods must be of metal or
metallized and connected to the shield at the cable clamp. Filter D-sub, especially J2, improves the immunity against
strong radio frequency signals. Heavy electrostatic discharge may happen, when weighing big containers, trucks,
plastic material or cases, especially when moving and in winter time. In this case, a separate ground cable, 2.5 to
10mm , must be drawn parallel with the transducer cable to a screw on the indicator case. 

2

All external connections must be made in separate grounded metal boxes.
Beware of high temperature sources, as the signal is very sensitive to temperature gradients.
The indicator must be well protected against high humidity, dust and dirt. Avoid to touch the boards.
W hen the displayed weight is unstable, always check connections, especially shield and ground.

W eight change (flow) function.

The weight change with respect to an external signal or during an internal time, selectable between 0.1 and 1000
seconds, may be measured with U137 and 237 series indicators.

Cs13:=1  Change with respect to a signal on the CTS input, minimum 100ms long.
Cs13:=2  Change every measurement cycle.
Cs13:=3  Change during 100 measurement cycles.

If 

)

W  is the displayed weight change, the average weight change per second is:

 Cs13:=2

 Cs13:=3

 Cycle s

 Cs08:1
 Cs08:4
 Cs08:7
 Cs08:1
 Cs08:4
 Cs08:7

 10

)

W

 

)

W

 

)

W /10

 

)

W /10

 

)

W /100

 

)

W /1000

 0.1
 1
 10
 10
 100
 1000

{F} {NET/GROSS} switches between gross weight mode and weight change mode.
The weight change mode is indicated by the COUNT indicator.
Count mode does not work.
The setpoints are compared with the displayed weight change, if Cs14:0 is chosen.
No mean value is calculated.

Pictures:

U1370

  

U2370       U2371

 U2372

  U2373

U2375       U2379 

Содержание U137 Series

Страница 1: ...21mV V W6 cut off There is a 3 pole filter with low ringing The response time is 1 6 seconds or 0 5 seconds with S2 1 2 and 3 on Keyboard functions at calibration The software calibration is performed...

Страница 2: ...RS422 or in Cs1 3 4 RS422 is addressed by ASCII 124 1 to 14 CR No one is addressed by and character other than 1 to 9 CR is recommended 15 Always addressed Used for single RS4225 unit without address...

Страница 3: ...3 4 ranges 1 at 10 20 50 2 at 10 25 50 5 at 10 20 40 4 5 ranges 1 at 5 10 25 50 2 at 4 10 20 40 5 at 5 10 20 50 5 6 ranges 1 at 2 4 10 20 40 2 at 2 5 10 20 50 5 at 2 5 5 10 25 50 6 Not used 7 Calibrat...

Страница 4: ...change in AD value 25 Third calibration point 00000 Display data 3 Must be even number 000000 Internal AD value Cs25 is not used when this value is lower than that in Cs24 Full AD converter range is...

Страница 5: ...both tilts and used in the formula The calibration can not be left with S1 when displaying angle These errors are caused by different angle between tilt sensor and load cell s and by the side force se...

Страница 6: ...mm month and dd day Option 190 Time as hh mm ss where hh are hours mm minutes and ss seconds Option 191 Date as 20yy mm dd where yy are the two last digits of the year mm month and dd day Option 192 W...

Страница 7: ...tation voltage 10V 1 However U2379 5V 1 Connect GND to metal hood of D sub J3 U237 series Power interface connector 9p D sub male B01910 1 2 3 4 5 6 7 8 9 0V Return for J3 2 Power V 0VD VB V after fus...

Страница 8: ...25mA SYNC CK SYNC D1 SYNC D2 TARE PT RET PRINT TEST DTR Connect to metal hood of D sub RS232 B01910 U137 has inverted TTL output Use J3 for RS232 RS232 U137 has inverted TTL output Use J3 for RS232 R...

Страница 9: ...be made in separate grounded metal boxes Beware of high temperature sources as the signal is very sensitive to temperature gradients The indicator must be well protected against high humidity dust and...

Страница 10: ...ht Indicators Date 141027 B01950 10 10 Unisystem AB Bergebyv 24 Tel 0560 140 55 www unisystem se 685 34 TORSBY S te Fax 0560 101 25 info unisystem se Old display U2370 U2371 U2372 U2373 U2375 U2379 U2...

Отзывы: