background image

 

 

NOTE: Wrong polarization may destroy the sensor  
 
For most sensors, the polarization will never need to be modified. The N

2

O sensor requires that the 

polarization is adjusted manually as this requires a period of pre-polarization at a more negative 
polarization before it is set to the measuring polarization (see the Nitrous Oxide Sensor Manual. To 
be downloaded at: 

https://www.unisense.com/manuals/

).  

 

7.4

 

Adjusting the offset 

The Offset may be set manually. It may be used to displace the signal in a positive or negative 
direction (Figure 3).  
 

7.5

 

Setting STOX sensor parameters  

The STOX sensor is a double O

2

 sensor, one sensor acts as the measuring sensor and the other acts 

as the front guard. The front guard will, when it is polarized, prevent O

2

 from reaching the measuring 

O

2

 sensor. The measuring signal, with the front guard on, is thus the signal for zero oxygen. The 

measuring signal, with the front guard off, is thus the signal for O

2

. This difference in signal with 

front guard on and off is proportional to the O

2

 concentration. Please see the STOX Sensor Manual 

for details (

https://www.unisense.com/manuals/

). 

 
The cyclic polarization of the front guard must be set individually for each sensor depending on the 
sensor characteristics. See the STOX sensor manual for details.  
 
When a STOX sensor is connected to a pA channel, the 

Enable STOX-mode

 checkbox will be available 

(Figure 4). Check this box to display the 

STOX zero on-time (s) 

and

 STOX zero off-time (s)

 fields.  

 
Setting the cyclic front guard polarization timing (seconds): 

 

The front guard on period is specified in the 

STOX zero on-time (s)

 field 

 

The front guard off period is specified in the 

STOX zero off-time (s)

 field 

Содержание UniAmp fx-3

Страница 1: ...Unisense UniAmp Multichannel Systems User Manual UNIAMP MULTICHANNEL SYSTEM MANUAL Copyright Unisense A S Version 1 1 August 2020...

Страница 2: ...ting the sensor polarization 6 7 4 Adjusting the offset 7 7 5 Setting STOX sensor parameters 7 7 6 Polarizing amperometric sensors without a PC connected 8 8 Sensor connections and compatibility 8 8 1...

Страница 3: ...ation For further information please see the document General terms of sale and delivery of Unisense A S as well as the manuals for the respective products 3 Support ordering and contact information I...

Страница 4: ...fication Area also called System tray in the lower right hand area of the screen Using this pop up you can adjust channel settings such as pre amplification offset sensor polarization etc 5 Operating...

Страница 5: ...e drivers needed for connection will automatically be installed The Unisense UniAmp service is installed together with the SensorTrace Suite Click on the UniAmp service icon to open a window with a fr...

Страница 6: ...sor signal in pico ampere pA is converted into a signal in millivolt mV This is set by the E2 PROM in the sensor when the sensor is plugged in but can be changed manually Simply select the Pre Amp Ran...

Страница 7: ...revent O2 from reaching the measuring O2 sensor The measuring signal with the front guard on is thus the signal for zero oxygen The measuring signal with the front guard off is thus the signal for O2...

Страница 8: ...or Manual To be downloaded at https www unisense com manuals 8 Sensor connections and compatibility 8 1 Sensors connectors and adaptors Sensors for the UniAmp meters must have the 10 pin UniAmp connec...

Страница 9: ...is used in the liquid 8 3 Software The UniAmp instruments require the latest version of the SensorTrace Suite software Check the version of your software in any of the applications Logger Profile Rat...

Страница 10: ...ure that you enter manually The input from a temperature sensor This sensor must be selected from the dropdown menu on the calibration window for each amperometric sensor The temperature correction of...

Страница 11: ...Update Microsoft Windows Update may interfere with logging for longer periods of time Windows may force installation of updates and restart the PC This will obviously stop any logging of data in Sens...

Страница 12: ...et and that it is not running on battery In the power saving settings on you PC disable all power saving options Please note that it may be necessary to disable power saving options both in Windows se...

Страница 13: ...this type will often be periodic e g noise emitted when equipment is switching on or off If such problems occur try switching nearby equipment off one by one to identify which equipment is generating...

Отзывы: