The standard multichannel UniAmp come in the following configurations: 

 

fx-6 UniAmp (2 x pA, 2 x mV, 1 x Op, 1 x T) 

 

x-5 UniAmp (2 x pA, 2 x mV, 1 x T) 

 

fx-3 UniAmp (1 x pA or 1 x mV, 1 x Op, 1 x T) 

 

 

Figure 1: fx-6 UniAmp 

All amperometric channels (pA) have support for the ultrasensitive STOX oxygen sensor with a 
detection limit of <10 nM.  
 
The fx6 and x-5 UniAmp are also available in customized versions, with the combination of channels 
decided by the customer.   
 
The meters offer temperature compensation for all amperometric channels with a range of ± 3°C 
from the calibration temperature. This will work if a temperature sensor is connected.   
 
The UniAmp meters are USB-powered and fully integrated with the Unisense SensorTrace Suite 
software, allowing any combination of Unisense electrochemical sensors and optode technology.  
 
The user interface consists of a pop-up window (a Microsoft Windows service) that is installed in the 

Windows Notification Area

, also called 

System tray

, in the lower right hand area of the screen. Using 

this pop-up, you can adjust channel settings such as pre-amplification, offset, sensor polarization etc.  

5

 

Operating and setup conditions 

The UniAmp meters are designed for laboratory use. Care must be taken to protect the amplifiers in 
humid, dusty, and corrosive environments (e.g. field conditions) since it may affect the performance, 
cause damage, or shorten lifetime. If the amplifier is used outdoors, it should be protected in a 
plastic bag or similar.  
 
Temperature:   0 to +50°C (<90% rel. humidity, non-condensing) 
Power:   

From PC, through the USB cable 

Grounding:  

Through the USB cable. Ensure that the PC is properly grounded.  

6

 

Installation and setup of UniAmp and software 

The UniAmp instruments must be used with the SensorTrace Suite software. Install the latest version 
of the software, which can be downloaded from 

https://www.unisense.com/Software_download/

The Logger and Programming Tool parts of the software suite are freeware, whereas Profile, Rate, 
and Photo require a license. The SensorTrace Suite is compatible with Windows versions 7 and 10.  
 

Содержание UniAmp fx-3

Страница 1: ...Unisense UniAmp Multichannel Systems User Manual UNIAMP MULTICHANNEL SYSTEM MANUAL Copyright Unisense A S Version 1 1 August 2020...

Страница 2: ...ting the sensor polarization 6 7 4 Adjusting the offset 7 7 5 Setting STOX sensor parameters 7 7 6 Polarizing amperometric sensors without a PC connected 8 8 Sensor connections and compatibility 8 8 1...

Страница 3: ...ation For further information please see the document General terms of sale and delivery of Unisense A S as well as the manuals for the respective products 3 Support ordering and contact information I...

Страница 4: ...fication Area also called System tray in the lower right hand area of the screen Using this pop up you can adjust channel settings such as pre amplification offset sensor polarization etc 5 Operating...

Страница 5: ...e drivers needed for connection will automatically be installed The Unisense UniAmp service is installed together with the SensorTrace Suite Click on the UniAmp service icon to open a window with a fr...

Страница 6: ...sor signal in pico ampere pA is converted into a signal in millivolt mV This is set by the E2 PROM in the sensor when the sensor is plugged in but can be changed manually Simply select the Pre Amp Ran...

Страница 7: ...revent O2 from reaching the measuring O2 sensor The measuring signal with the front guard on is thus the signal for zero oxygen The measuring signal with the front guard off is thus the signal for O2...

Страница 8: ...or Manual To be downloaded at https www unisense com manuals 8 Sensor connections and compatibility 8 1 Sensors connectors and adaptors Sensors for the UniAmp meters must have the 10 pin UniAmp connec...

Страница 9: ...is used in the liquid 8 3 Software The UniAmp instruments require the latest version of the SensorTrace Suite software Check the version of your software in any of the applications Logger Profile Rat...

Страница 10: ...ure that you enter manually The input from a temperature sensor This sensor must be selected from the dropdown menu on the calibration window for each amperometric sensor The temperature correction of...

Страница 11: ...Update Microsoft Windows Update may interfere with logging for longer periods of time Windows may force installation of updates and restart the PC This will obviously stop any logging of data in Sens...

Страница 12: ...et and that it is not running on battery In the power saving settings on you PC disable all power saving options Please note that it may be necessary to disable power saving options both in Windows se...

Страница 13: ...this type will often be periodic e g noise emitted when equipment is switching on or off If such problems occur try switching nearby equipment off one by one to identify which equipment is generating...

Отзывы: