Updates are important for keeping Windows secure. Therefore, if you choose to disable or delay 
updates, make sure to allow Windows to check for updates now and then. 
 

13.1.2

 

Loss of connection to the UniAmp because of USB power saving 

Windows may turn USB connections off in order to save power. To avoid this, open 

Device manager

double click the 

Ports (COM & LPT)

 to unfold it, double click 

UniAmp

 to open the 

UniAmp Properties

 

window (Figure 7). Make sure that the checkbox 

Allow the computer to turn off this device to save 

power

 is unchecked.  

 

 

Figure 7: Power management of the USB connection to the UniAmp. 

13.1.3

 

Disable power saving on the PC 

Make sure that the PC is powered from a stable mains outlet and that it is not running on battery. In 
the power saving settings on you PC, disable all power saving options. Please note that it may be 
necessary to disable power saving options both in 

Windows settings

 as well as in any proprietary 

power management software on the PC.  
 

13.2

 

Electrical noise  

If the signal is noisy, it may be caused by electrical interference. This may be solved by ensuring 
proper grounding of the whole measuring setup: 

 

The cable between the mains supply and the power supply of the PC must be grounded 

 

The mains supply must be properly grounded 

 

Connect the ground connector of the UniAmp to the liquid where the sensor is measuring, 
using the grounding cable that came with the UniAmp. Connect the grounding cable banana 
plug to the connector on the back side of the UniAmp (Figure 5) and dip the other end into 
the liquid where the sensor is measuring (see also 

8.2 Grounding

) 

 

Содержание UniAmp fx-3

Страница 1: ...Unisense UniAmp Multichannel Systems User Manual UNIAMP MULTICHANNEL SYSTEM MANUAL Copyright Unisense A S Version 1 1 August 2020...

Страница 2: ...ting the sensor polarization 6 7 4 Adjusting the offset 7 7 5 Setting STOX sensor parameters 7 7 6 Polarizing amperometric sensors without a PC connected 8 8 Sensor connections and compatibility 8 8 1...

Страница 3: ...ation For further information please see the document General terms of sale and delivery of Unisense A S as well as the manuals for the respective products 3 Support ordering and contact information I...

Страница 4: ...fication Area also called System tray in the lower right hand area of the screen Using this pop up you can adjust channel settings such as pre amplification offset sensor polarization etc 5 Operating...

Страница 5: ...e drivers needed for connection will automatically be installed The Unisense UniAmp service is installed together with the SensorTrace Suite Click on the UniAmp service icon to open a window with a fr...

Страница 6: ...sor signal in pico ampere pA is converted into a signal in millivolt mV This is set by the E2 PROM in the sensor when the sensor is plugged in but can be changed manually Simply select the Pre Amp Ran...

Страница 7: ...revent O2 from reaching the measuring O2 sensor The measuring signal with the front guard on is thus the signal for zero oxygen The measuring signal with the front guard off is thus the signal for O2...

Страница 8: ...or Manual To be downloaded at https www unisense com manuals 8 Sensor connections and compatibility 8 1 Sensors connectors and adaptors Sensors for the UniAmp meters must have the 10 pin UniAmp connec...

Страница 9: ...is used in the liquid 8 3 Software The UniAmp instruments require the latest version of the SensorTrace Suite software Check the version of your software in any of the applications Logger Profile Rat...

Страница 10: ...ure that you enter manually The input from a temperature sensor This sensor must be selected from the dropdown menu on the calibration window for each amperometric sensor The temperature correction of...

Страница 11: ...Update Microsoft Windows Update may interfere with logging for longer periods of time Windows may force installation of updates and restart the PC This will obviously stop any logging of data in Sens...

Страница 12: ...et and that it is not running on battery In the power saving settings on you PC disable all power saving options Please note that it may be necessary to disable power saving options both in Windows se...

Страница 13: ...this type will often be periodic e g noise emitted when equipment is switching on or off If such problems occur try switching nearby equipment off one by one to identify which equipment is generating...

Отзывы: