Page 19
Choose the Sensor you wish to calibrate. The sensor name, type and calibration unit are
shown for each sensor.
The Sensor signal displays the measured raw sensor signal for the selected sensor. The
sensor signal can also be followed continuously in the Live Data graph at the bottom of the
screen.
7.6.2
Calibration of recorded data
The FDL Mini only stores raw data. The calibration data are stored in a Logger file on a PC
and are applied to the FDL Mini datafile after logging is completed.
Open the data file in Logger using
File - Open
dialog. Make the dataset active by double-
clicking the appropriate experiment name in the
Experiment overview
column. The
uncalibrated data will now be shown in a new tab.
Import the calibration by opening the Logger file containing the calibration data using the
File - Import calibration
dialog.
Select a sensor in the green column to the left and click the radio button it the
In use
column
to select the appropriate calibration. Click OK in the pop-up box to apply the selected
calibration. Repeat this for each sensor.