Page 14
Select
Channel settings
and in the window, that opens, you can set the polarization. If the
Enable STOX-mode
is selected the STOX-timing will be shown and can be changed (see the
manual for the STOX sensor for selecting the correct on- and off-time and polarization).
NOTE: If selecting or deselecting
Enable STOX-mode
, the grey
Scan
button in the upper left
of the main Logger window must be clicked to apply the changes to the FDL Mini. This will
create a new logging file in the FDL Mini.
7.3.2
Setting sampling rate
Right click one of the channels and select
Device settings
. In the window that opens you may
select the datalogging frequency for the FDL Mini. Logging in the FDL Mini may occur at the
following intervals: 1 sec. (Middle), 10 sec. (Low) or 60 sec. (Very low). Data from all
channels are logged at the same frequency.
NOTE: The
Sampling rate
settings only affect the internal data logging on the FDL Mini.
When the FDL Mini is connected to a PC the data transfer frequency is always 1 Hz but the
internal logging in the FDL Mini still runs at the frequency specified in the
Sampling rate
window.