7.3.5
Setting channel information
Sensor:
Mark the checkboxes for the channels/sensors you want to view and record signals
from.
Type:
Choose sensor type from the drop-down menu if the default type is not correct.
Unit:
Select an appropriate concentration unit for the sensor signal when calibrated.
Range (V):
Select the voltage range for the amplifier. Select the smallest range possible to
get the most out of the resolution of the amplifier matching the expected signal range of the
sensors. It is recommended not to select an unnecessarily high range either, as this may
cause loss of resolution. However, the range should not be chosen so small that the signal
gets beyond the selected range as this will cause the amplifier to get out of range.
Name:
Write a name describing your sensor (optional).
Comment:
Write a comment about your sensor (optional).