Calibration
33
Calibr
a
ti
on
Chapter
4
Input procedure of rated capacity
Decimal place and enabled number of digits of torque are determined according to the input value
of the rated capacity.
The possible number of digits for setting is up to five in total and up to three after the decimal point.
e.g.) When UTM
Ⅱ
-10Nm is used
i) When rated capacity is 10Nm
Torque readings
-10, -9, ... 0, 1, ... 9, 10
ii) When rated capacity is 10.0Nm
Torque readings
-10.0, -9.9, ... 0.0, 0.1, ... 9.9, 10.0
iii) When rated capacity is 10.00Nm
Torque readings
-10.00, -9.99, ... 0.00, 0.01, ... 9.99, 10.00
iv) When rated capacity is 10.000Nm
Torque readings
-10.000, -9.999, ... 0.000, 0.001, ... 9.999, 10.000
v) When rated capacity is 10.0000Nm
Four-digit decimal point cannot be set.
The following error message appears when the calibration is performed, and the rated capacity
is rewritten to the settable digits.
Negative value can be input to the rated capacity.
At this point, the sign of the torque readings can be inverted.
Key points