31
4
Calibration
31
Calibr
a
ti
on
Chapter
4
(6) a. Equivalent input calibration
Rated output and rated capacity of UTM
Ⅱ
are registered.
<Setting value>
Rated output [V]:
-6.000 to 6.000 (excluding 0)
Display value (Rated capacity):
In case of zero-digit decimal point:
-99999 to 99999 (excluding 0)
In case of one-digit decimal point:
-9999.9 to 9999.9 (excluding 0.0)
In case of two-digit decimal point:
-999.99 to 999.99 (excluding 0.00)
In case of three-digit decimal point:
-99.999 to 99.999 (excluding 0.000)
■
Execution procedure
1.
Input the rated output and rated capacity.
(The rated capacity unit depends on the
setting of (2).)
Refer to P.33 "Input procedure of rated capacity" for input procedure.
2.
Select "Equivalent" for calibration mode.
3.
Click the "Execute" button.
4.
When the "Execute" button is
clicked, a confirmation dialog to
overwrite is displayed, and click "Yes
(Y)".
Set the display value (rated capacity) that has been converted according to the
set torque unit and decimal place.
Refer to P.37 "4-6.Unit conversion table" for unit conversion table.
Key points