36
5
EXPLANATION OF INDICATED- VALUE-RELATED FUNCTIONS
36
EXPLANATIO
N OF INDI
CATED-
VAL
U
E-REL
A
TED FUNCT
IONS
Chapter
5
5-2.
Analog filter
This is a low-pass filter for filtering input signals from the strain gauge sensor, and canceling
unwanted noise components (a filter that passes only the signals with a specific frequency or less,
and attenuates signals with higher frequencies). With an increase in the cut-off frequency, the
response becomes faster, but noise components may be displayed. It can be set for each of the ch1
and ch2 inputs.
≪
Setting value
≫
0
:
3Hz
1
:
30Hz
2
:
300Hz
3
:
1kHz
■
Analog filter setting method
1.
Select setting mode 3.
Press
→
twice.
2.
Set "Analog filter."
Press
twice.
⇒
" Analog filter " is displayed.
Press
, and present set value blinks.
Input the setting value with
.
Press
to validate it.
(
Initial value: 2
)
3.
Press
twice to go back to the indicated value display.
Ch1 Ch
2
H I
LO
H I
LO
FNC
H I
Ch1 Ch
2
OK
LO
PEAK
HOLD
H I
LO
H I
LO