Technical data
1.3 Data and specifications
Gas inputs
Input process gas:
1
Output process gas:
1
Output condensate:
1
Gas connections:
Compression fitting 6mm / ¼”
Max. gas input pressure: 750 mbar absolute (11 psi)
Min. gas input pressure:
-100 mbar relative
Gas pressure drop:
5 - 10 mbar (2 - 4” H2O)
Relative gas humidity:
Performance data
< 95% (non-condensing)
Gas capacity, max
250 l/h
Effective cooling capacity: 20 W at
∆
T 25 °C
Gas temperature difference,
max:
Max. 25 K
Protection class:
I
Degree of protection:
Voltage supply
IP 22
Voltage:
100 - 240 VAC, 50 - 60 Hz
Power consumption:
250 VA max.
Protection class:
I
Degree of protection:
Interfaces
IP22
Relay Contacts:
Environmental conditions
2
Operating temperature:
5 - 45 °C. (40 - 115°F)
Humidity:
0 - 95 % relative humidity
Ambient pressure:
0.8 - 1.1 bar (11.6 - 16 psi)
Storage temperature:
Weight
-20 - 60 °C ( -4 - 140°F)
Weight:
Approx. 12.6 kg. (28lbs)
7
Содержание Pury250 S
Страница 1: ...Operating Instructions Gas Cooler Pury250 S ...
Страница 5: ...1 Technical data 1 1 Dimensions Weight approx 12 6 kg 28lbs 5 ...
Страница 10: ...10 EC Declaration of Conformity ...
Страница 15: ...15 4 Protective equipment 4 1 Enclosure cover Fig 4 1 Enclosure cover ...
Страница 32: ...32 Transport installation and acceptance ...
Страница 34: ...34 Commissioning Switching on ...
Страница 36: ...36 Description of the HMI and operator control elements ...
Страница 49: ...49 Maintenance ...
Страница 53: ...53 ...
Страница 54: ...54 ...
Страница 58: ...58 Service ...
Страница 59: ...59 14 Related documents Declaration of Conformity Service documentation optional ...
Страница 60: ...60 Related documents ...
Страница 62: ...62 Disposal ...
Страница 64: ...64 Spare parts ...