42
Operation
9.3.4 Difference between ambient and cooler temperatures
Difference is a measure for the degree of drying. Default value is 10°C, setting
range is "Off", 10°C to 30°C.
"Off" means that no temperature difference is permitted. Cooling to the dew point
temperature will be carried out.
9.3.5 Pump condensate quantity
Information on the expected condensate quantity, pump delivers this quantity
automatically at intervals.
Default value is 5 ml/h, setting range is 5 ml/h to 150 ml/h.
Note: The service life of the condensate pump is dependent on the flow rate.
Adjust the condensate quantity to the conditions.
Table 1: Condensate quantity
Condensate quantity for: Dew point 5°C and 250 liter gas saturated at 100 mbar
gauge pressure.
,
0
00
00
,
5
00
,
10
00
,
15
,
00
20
25
00
,
00
,
30
35
00
,
00
,
40
5
10
15
20
25
30
40
45
50
55
60
35
gas temperature
condensate ml/h
Содержание Pury250 S
Страница 1: ...Operating Instructions Gas Cooler Pury250 S ...
Страница 5: ...1 Technical data 1 1 Dimensions Weight approx 12 6 kg 28lbs 5 ...
Страница 10: ...10 EC Declaration of Conformity ...
Страница 15: ...15 4 Protective equipment 4 1 Enclosure cover Fig 4 1 Enclosure cover ...
Страница 32: ...32 Transport installation and acceptance ...
Страница 34: ...34 Commissioning Switching on ...
Страница 36: ...36 Description of the HMI and operator control elements ...
Страница 49: ...49 Maintenance ...
Страница 53: ...53 ...
Страница 54: ...54 ...
Страница 58: ...58 Service ...
Страница 59: ...59 14 Related documents Declaration of Conformity Service documentation optional ...
Страница 60: ...60 Related documents ...
Страница 62: ...62 Disposal ...
Страница 64: ...64 Spare parts ...