![Union Instruments Pury250 S Скачать руководство пользователя страница 50](http://html.mh-extra.com/html/union-instruments/pury250-s/pury250-s_operating-instructions-manual_843042050.webp)
50
Maintenance
11.2.1 Replacing filter inserts
Disassembly
To replace the filter inserts, the cooler tube must be removed.
1. Unscrew compression fittings of the input process gas and output process
gas.
2. Unscrew compression fitting of condensate hose on cooler tube.
3. Screw off lower part from the tube.
4. Pull out tube upwards from the cooler.
5. Screw off upper part from the tube.
Fig. 11.1:
Disassembly of cooler tube
2
1
3
4
5
1