Operation
68
Main Menu - Options - I/O - Digital Outputs
Digital signals (relay outputs and floating change-over contacts) are configured in
this menu.
The following must be observed for this:
Zero position:
The zero position of the digital outputs can be freely selected: low/high.
Process:
The combustion gas solenoid valve is open, and the flame is burning.
Calibration:
The calibration gas solenoid valve is open, and the flame is burning.
Maintenance:
The device must be checked in the foreseeable future, but is still ready for
operation. The reason for maintenance can be obtained from the “General
information” field.
Examples reasons include:
•
Filter change (air filter)
•
No calibration gas (calibration canceled)
•
Room temperature too high (air inlet temperature > 52°C)
•
Gas pressure is too low
•
Calibration deviation outside of tolerances
•
The calibration was unstable and was canceled
Содержание CWD3000 EXP
Страница 6: ...Technical data 6 1 2 Dimensions Fig 1 2 Housing dimensions Weight approx 140 kg ...
Страница 14: ...EU Declaration of Conformity 14 ...
Страница 30: ...Description and connections 30 ...
Страница 56: ...Description of the work stations operator control elements 56 ...
Страница 62: ...Operation 62 ...
Страница 86: ...Operation 86 Main menu Options System Settings Change color The display colors can be changed here ...
Страница 95: ...Operation 95 Show startup Selection of the start events Show calibration Selection of the calibration events ...
Страница 98: ...Decommissioning Switching off 98 ...
Страница 103: ...Maintenance 103 ...
Страница 104: ...Maintenance 104 ...
Страница 110: ...Troubleshooting 110 ...
Страница 112: ...Service 112 ...
Страница 114: ...Disposal 114 ...
Страница 116: ...Spare parts 116 ...