Technical data
12
1.14 Ambient conditions of installation location
NOTICE
Use of the combustion calorimeter outside the approved range of ambient
conditions results in voiding of the approval! Risk of serious accidents.
Installation location:
Protect against direct sunlight and ensure sufficient
shade.
Temperature change:
≤ 5°C per hour
Air humidity:
0 – 95% rel. humidity
Ambient pressure:
800 – 1,100 hPa (0.8 – 1.1 bar)
Direct sunlight can cause changes in temperature; temperature changes above
5°C/h affect measurement results.
For other requirements, see Section 6.3.