Troubleshooting
108
12.5.2 Error/status messages
Int. pressure
•
Alarm in the event of inadequate combustion gas or calibration gas.
•
Adjustable threshold; default: 14 mbar
Overtemp
•
The flame is too hot. Wobbe signal > 76 mV
•
Temperature rise in the inner tube of the thermoelectric battery > 50°C
Fault
•
Gas pressure is too low.
•
Fault in the thermoelectric battery.
•
Defective PT 100 temperature sensor.
Service
•
Air inlet temperature > 52°C
Operation
•
Flame is burning, and calorimeter is running on combustion or calibration gas.
Operation delay
•
Flame is burning, calorimeter is running on combustion or calibration gas.
•
Delay time has elapsed, after which the analog signals are enabled.
Process
•
Flame is burning, and solenoid valve for calibration gas is open.
•
Calorimeter is running on combustion gas.
Содержание CWD3000 EXP
Страница 6: ...Technical data 6 1 2 Dimensions Fig 1 2 Housing dimensions Weight approx 140 kg ...
Страница 14: ...EU Declaration of Conformity 14 ...
Страница 30: ...Description and connections 30 ...
Страница 56: ...Description of the work stations operator control elements 56 ...
Страница 62: ...Operation 62 ...
Страница 86: ...Operation 86 Main menu Options System Settings Change color The display colors can be changed here ...
Страница 95: ...Operation 95 Show startup Selection of the start events Show calibration Selection of the calibration events ...
Страница 98: ...Decommissioning Switching off 98 ...
Страница 103: ...Maintenance 103 ...
Страница 104: ...Maintenance 104 ...
Страница 110: ...Troubleshooting 110 ...
Страница 112: ...Service 112 ...
Страница 114: ...Disposal 114 ...
Страница 116: ...Spare parts 116 ...