![Union Instruments CWD3000 EXP Скачать руководство пользователя страница 37](http://html.mh-extra.com/html/union-instruments/cwd3000-exp/cwd3000-exp_operating-instructions-manual_843047037.webp)
Transport, installation and acceptance
37
6.3.3 Gas connections
DANGER
Risk of explosion!
Non-approved lines and insufficient connections pose a hazard!
Gas connections may only be installed by qualified personnel!
Check for proper sealing of connections and ensure this during operation!
NOTE
•
Accessories must be clean and free of residue. Contaminants may reach
inside the combustion calorimeter and may cause false measurements
and/or damage.
•
The input pressure for the gas connections must not exceed the
specifications of the technical data of the combustion calorimeter.
•
Check each connection carefully for leaks. When leaks are present, the
system draws air and indicates false measured values.
•
Do not use sealant for sealing the gas connections. Sealant components
may falsify the measurement result.
•
Only suitable lines may be used.
6.3.3.1 Ignition protection gas supply – Compressed air
Compressed air is required as the ignition protection gas for operation of the
combustion calorimeter. Ensure sufficient supply and seal of the connections –
check connection point for proper seal.
Device requires compressed air at 4.5 – 5 bar, 16,000 – 30,000 l/h, for operation.
Air recommendation as per ISO 8573-
1, class 533 (solid material: 40 μm (class
5)/dew point: –20°C (class 3)/oil quality: 1 mg/m³ (class 3))
NOTE
•
When installing a shut-off valve of the ignition protection gas supply
Post warning information: Warning – Valve of the ignition protection gas
supply – Follow instructions before closing!
Содержание CWD3000 EXP
Страница 6: ...Technical data 6 1 2 Dimensions Fig 1 2 Housing dimensions Weight approx 140 kg ...
Страница 14: ...EU Declaration of Conformity 14 ...
Страница 30: ...Description and connections 30 ...
Страница 56: ...Description of the work stations operator control elements 56 ...
Страница 62: ...Operation 62 ...
Страница 86: ...Operation 86 Main menu Options System Settings Change color The display colors can be changed here ...
Страница 95: ...Operation 95 Show startup Selection of the start events Show calibration Selection of the calibration events ...
Страница 98: ...Decommissioning Switching off 98 ...
Страница 103: ...Maintenance 103 ...
Страница 104: ...Maintenance 104 ...
Страница 110: ...Troubleshooting 110 ...
Страница 112: ...Service 112 ...
Страница 114: ...Disposal 114 ...
Страница 116: ...Spare parts 116 ...