7.2.3 Messages/malfunctions
Display of messages/malfunctions
Pressure switch triggers, required overpressure not reached
•
Check filter of compressed air supply for free flow / dirt
•
Check inlet pressure of compressed air supply, see technical data
•
Check function of the pressure switch
•
Check function of the faceplate in the exhaust air duct system
•
Check free flow of compressed air in all concerned lines/connections