Страница 1: ...1 C6 TRMS Digital Clamp Meter 600AAC V1 0 29 11 2022...

Страница 2: ...2...

Страница 3: ...niks it The registration of your products will allow you to stay informed about news take advantage of advantageous discounts dedicated to you for the purchase of accessories and products for your dai...

Страница 4: ...11 3 5 Resistance measurement 12 3 6 Diode measurement 13 3 7 Circuit continuity measurement 13 3 8 Capacitance measurement 13 3 9 Frequency and duty ratio measurement 14 3 10 Temp measurement 14 3 11...

Страница 5: ...sed circuit measurement It is a superior instrument special for electricians 1 1 Safety Standards This instrument is designed and manufactured according to the safety standards IEC61010 1 and IEC61010...

Страница 6: ...C or 30V AC be careful and remember to keep your fingers behind the hand shelter of the probe When the range of to be measured value is uncertain set the function range to the maximum range position D...

Страница 7: ...ument Schematic 1 AC current clamp Used for sampling AC current 2 Safety barrier Effectively keep the operator from accidentally taping the live conductor during the current measurement 3 Function ran...

Страница 8: ...off automatically after about 30 seconds 5 Inrush current test button Tap the INRUSH button to turn on or off the inrush current measurement function Press and hold the INRUSH button for more than 2...

Страница 9: ...es Press and hold the MIN MAX button for more than 2 seconds to exit the maximum and minimum value measurement 11 Function selection button In position it is used to switch between current value and c...

Страница 10: ...ent Do not apply voltage exceeding 600V DC or 600V AC RMS between the common terminal and the ground to prevent electric shock and or damage to the instrument 3 1 AC and DC voltage measurement Insert...

Страница 11: ...et the function range switch to the maximum range and then gradually reduce the range until a satisfactory resolution is achieved When measuring high voltage pay special attention to avoid electric sh...

Страница 12: ...probe into the COM terminal 2 Set the function range switch to position tap the SEL key to switch to function and connect the probe to the measured resistance 3 Read the measurement result after the r...

Страница 13: ...probe into the INPUT terminal and the black probe into the COM terminal 2 Set the function range switch to the position 3 Tap the SEL button to switch to position and the probe is connected to two po...

Страница 14: ...black probe into the COM terminal 2 Set the function range switch to the position and tap the SEL button to switch to display the frequency or duty ratio of the measured signal 3 Connect the probes to...

Страница 15: ...more than 20 seconds to avoid burning out caused by excessive heating of load resistance 3 12 Automatic Shutdown When there is no operation within 10 minutes after startup the instrument will enter th...

Страница 16: ...16 Caution When measuring AC current please place the conductor to be measured in the center of the clamp If not the position error of 1 5 to the maximum can be increased...

Страница 17: ...ltage Range Resolution Accuracy 6V 0 001V 0 8 reading 5 digits 60V 0 01V 600V 0 1V Input impedance 10M Measuring frequency range 40Hz 1000Hz Maximum allowable input voltage 600V DC or 600V AC RMS AC c...

Страница 18: ...C or 600V AC RMS Diode test Range Resolution Accuracy 0 001V Displays the approximate value of the forward voltage drop The open circuit voltage is about 3 0V Overload protection 600V DC or 600V AC RM...

Страница 19: ...ts 10mF 0 001mF 100mF 0 01 mF 5 0 reading 5 digits Overload protection 600V DC or 600V AC RMS Frequency Range Resolution Accuracy 10Hz 0 001Hz 0 5 reading 5 digits 100Hz 0 01Hz 1kHz 1Hz 10kHz 0 01kHz...

Страница 20: ...ge Resolution Accuracy 1 99 0 1 3 0 reading 2 digits Overload protection 600V DC or 600V AC RMS Temperature Range Resolution Accuracy 20 1000 4 1832 1 1 1 0 reading 2 digits Overload protection 600V D...

Страница 21: ...Polarity indication automatic indication indicates negative polarity Over range display 0L or 0L Sampling time about 3 times second about 10 times second for simulation strip Unit display display of f...

Страница 22: ...case To avoid electric shock caused by false readings replace the battery immediately when the instrument displays Turn off the power when the instrument is not in use and turn the function range swi...

Страница 23: ...ans that the battery is low the deviation of the measured value in this state is greater and needs to be replaced Please follow the steps below 1 Disconnecttheprobefromthecircuitundertest 2 Turnthefun...

Страница 24: ...For shipping only use the original packaging Any damage caused by the use of non original packing shall be charged to the customer The manufacturer accepts no responsibility for damage caused to peopl...

Страница 25: ...ery and wear of the cables and replace them if necessary If the instrument continues to manifest malfunctions check if the procedure of use of the same is in accordance with what is indicated in this...

Страница 26: ...26...

Страница 27: ...27...

Страница 28: ...28 Uniks Srl https www uniks it info uniks it Via Vittori 57 48018 Faenza RA Italy 0546 623002 0546 623691...