7 Unihedron Device Manager
The original factory calibration values were provided on a paper with the shipped unit. If you have lost this information,
please contact Unihedron to have the information emailed.
7.7 Report Interval tab
The Report Interval tab is used to show and set the report interval information for the selected SQM. Report interval
settings are used to control the SQM to send readings out at a regular rate (in seconds) and only if darker than a specified
threshold value. See section 8.6 on page 53 for more information about “Interval reporting”.
Figure 7.27: Report interval tab
Use the “R” button to temporarily set the value in to RAM for experimentation. The new value will be used instantly.
The RAM value is set to the EEPROM value on power-up.
Use the “E/R” button to set the value in to EEPROM and RAM. The new value will be used instantly and also after
power up.
Report interval is measured in seconds (i.e. 300 = 5 minutes).
Report threshold is measured in Magnitudes per square arcsecond (mpsas), a larger positive value is darker.
7.8 Firmware tab
In a case where new firmware is supplied by Unihedron to correct bugs or add features, use this tab to select and load
firmware in to the SQM-LU.
All available versions of firmware are shipped with the UDM software package in case you want to revert to an earlier
firmware version for testing purposes. New firmware versions are announced on the Unihedron forum.
Figure 7.28: Firmware tab
34
Unihedron SQM-LU Operator’s Manual -
20170728