41
D2
W
Cell #1
Photometric Validity Test
WL : 900.0nm %T:
12: 35: 27
No WL(nm) %T(Std) %T Result
1 500.0 9.28 9.28 pass
2 600.0 8.45 8.45 pass
3 700.0 5.66
F1:
Standard
F2:
Mode
F3:
Tolerance
Fig 94
J
WL Validity
Press
【
7
】
in Fig 73
to WL validity (Fig 95).
【
ESC/STOP
】
to exit.
D2
W
Cell #1
Wavelength Validity Test
WL : 900.0nm Abs:
12: 35: 27
No WL(nm) Peak(nm) T% Result
F1:
Set peaks
F2:
Mode
F3:
Tolerance
Fig 95
1. Press
【
F1
】
to set the standard peak. Press
【
ENTER
】
to edit and
input wavelength by pressing the numeric keypad (Fig96).
【
ESC/STOP
】
to finish inputting and exit.
D2
W
Cell #1
Wavelength Validity Test
WL : 900.0nm %T:
12: 35: 27
Input the standard:0
No WL(nm) Peak(nm) %T Result
1 241.4
2 361.0
3 417.0
4 537.6
5 641.4
6 807.4
Select
Fig 96
2. Press
【
F2
】
to select test mode (Abs or %T, Fig 97).