39
SPECTRO-QUEST
D2
W
Cell #1
1
2
3
9
8
4
6
7
WL Reset
Printer
Lamp
Clock
Accu Validity
WL Validity
Connect to PC
Beeper on/off
United Products & Instruments INC.
5
Dark current
F1
Delete entire saved files
F2
Restore default
WL : 656.1nm
Get dark current
UNICO SPECTROPHOTOMETER
Fig 88
J
Accu Validity
Press
【
4
】
In Fig73 to do
accu validity (Fig 89).
【
ESC/STOP
】
to exit.
D2
W
Cell #1
Photometric Validity Test
No WL(nm) Abs(Std) Abs Result
WL : 900.0nm Abs:
12: 35: 27
F1:
Standard
F2:
Mode
F3:
Tolerance
Fig 89
1.
Press
【
SET
λ】
to set the wavelength. Press
【
ENTER
】
to edit and
input wavelength by pressing the numeric keypad (Fig 90).
【
ESC/STOP
】
to finish inputting and exit.
D2
W
Cell #1
Photometric Validity Test
WL : 900.0nm Abs:
12: 35: 27
Please input wl:546
No WL(nm) Abs(Std) Abs Result
1 440.0
2
546.0
3 -more-
Fig 90
2.
Press
【
F1
】
to set the standard value, Press
【
ENTER
】
to edit and input
by pressing the numeric keypad (Fig 91).
【
ESC/STOP
】
to finish inputting
and exit.